Bador R, Déchaud H, Claustrat F, Desuzinges C
Clin Chem. 1987 Jan;33(1):48-51.
This time-resolved immunofluorometric assay for human follitropin involves use of europium- or samarium-labeled monoclonal antibodies, with an average incorporation ratio of 3 mol of Eu3+ or Sm3+ per mole of antibody. These lanthanide ions are bound to the antibody molecules by means of the anhydride of diethylenetriaminepentaacetic acid. The solid-phase antibody is immobilized inside polystyrene tubes in which plasma samples were assayed in a one-step procedure. After incubation, the fluorescence intensity of Eu3+ or Sm3+ label is measured by time-resolved fluorometry, with a nitrogen laser as the pulsed excitation source. The sensitivity of the assay is largely better with Eu3+ than with Sm3+ because of the difference in their intrinsic luminescence properties. Results obtained with the proposed methods correlated well with those by an immunoradiometric method.