University Hospital Jena, Department of Radiation Oncology, Germany; Institute for Medical Physics and Radiation Protection IMPS, University of Applied Science - THM, Giessen, Germany; Philipps-University, Marburg, Germany.
University Hospital Jena, Department of Radiation Oncology, Germany.
Z Med Phys. 2019 Dec;29(4):337-348. doi: 10.1016/j.zemedi.2019.03.003. Epub 2019 May 2.
Helical TomoTherapy allows a highly conformal dose distribution to complex target geometries with a good protection of organs at risk. However, the small field sizes associated with this method are a possible source of dosimetrical uncertainties. The IAEA has published detector-specific field output correction factors for static fields of the TomoTherapy in the TRS483. This work investigates the average subfield size of helical TomoTherapy plans.
A new parameter for helical TomoTherapy was defined - the fluence-weighted average subfield size. The subfield sizes were extracted from the leaf-opening time sinograms in the RT-plan files for 30 clinical prostate and head and neck plans and were put in relation to Delat4 Phantom+ measurement results. Additionally the influence of planning parameters on the subfield size was studied by varying the modulation factor, number of iterations and pitch in the dose optimization and calculation for three different clinical indications H&N, prostate and rectum cancer. Selected plans were dosimetrically verified by Delta4 measurements to examine the reliability in dependence of the average subfield size. Furthermore, the impact of the planning parameters on a) the dose distribution, with regard to the planning target volume and regions at risks, and b) machine characteristics such as delivery time, actual modulation factor and leaf-opening times were evaluated.
The average equivalent square subfield lengths (s¯) of the two investigated indications did not differ significantly - prostate plans: 2.75±0.14cm and H&N plans: 2.70±0.16cm, both with a jaw width of 2.5cm. No correlation between field size and measured dose deviation was detected. The number of iterations and the modulation factor have a considerable influence on the average subfield size. The higher the planned modulation factor and the more iterations are used during optimization, the smaller is the subfield size. In our pilot study plans were calculated with field sizes s¯ between 4.2cm and 1.7cm, with a jaw width of 2.5cm. Again, the measurement results of Delta4 showed no significant deviation from the doses calculated by the TomoTherapy planning system for the whole range of subfield sizes, and no significant correlation between field sizes and dose deviations was found. As expected, the clinical dose distribution improved with increasing modulation factor and an increasing number of iterations. The compromise between an improved dose distribution and smaller s¯ was shown.
In this work, a method was presented to determine the average subfield size for helical TomoTherapy plans. The response of the Delta4 did not show any dependence on field size in the range of the field sizes covered by the studied plans. The influence of the subfield sizes on other dosimetry systems remains to be investigated.
螺旋断层放疗(TomoTherapy)允许对复杂靶区进行高度适形的剂量分布,并能很好地保护危及器官。然而,与该方法相关的小射野尺寸可能是剂量学不确定性的一个来源。国际原子能机构(IAEA)已经在 TRS483 中发布了 TomoTherapy 静态场的特定探测器的射野输出校正因子。本研究调查了螺旋断层放疗计划的平均子野大小。
为螺旋断层放疗定义了一个新参数 - 通量加权平均子野大小。从 30 个临床前列腺和头颈部计划的射野开启时间正弦图中提取子野大小,并将其与 Delat4 Phantom+ 测量结果进行比较。此外,通过改变调制因子、迭代次数和螺距,在三种不同的临床适应证(头颈部、前列腺和直肠癌)的剂量优化和计算中,研究了规划参数对子野大小的影响。选择的计划通过 Delta4 测量进行了剂量验证,以检查平均子野大小的依赖性。此外,还评估了规划参数对 a)剂量分布的影响,包括计划靶区和危及器官区域,以及 b)机器特性的影响,如输送时间、实际调制因子和叶片开启时间。
两种研究适应证的平均等效正方形子野长度(s¯)没有显著差异 - 前列腺计划:2.75±0.14cm,头颈部计划:2.70±0.16cm,两者的颌宽均为 2.5cm。未检测到射野大小与测量剂量偏差之间的相关性。迭代次数和调制因子对平均子野大小有很大的影响。计划调制因子越高,优化过程中使用的迭代次数越多,子野尺寸就越小。在我们的初步研究中,计划的子野大小在 4.2cm 到 1.7cm 之间,颌宽为 2.5cm。同样,Delta4 的测量结果显示,在整个子野大小范围内,与 TomoTherapy 计划系统计算的剂量没有显著偏差,也没有发现射野大小与剂量偏差之间有显著相关性。正如预期的那样,随着调制因子和迭代次数的增加,临床剂量分布得到了改善。显示了在改善剂量分布和更小的 s¯之间的折衷。
在这项工作中,提出了一种用于确定螺旋断层放疗计划平均子野大小的方法。Delta4 的响应在研究计划所涵盖的射野尺寸范围内没有显示出任何依赖于射野尺寸的情况。子野大小对其他剂量测量系统的影响仍有待研究。