Kado Yuichiro, Byram Nicole, Miyamoto Takuma, Horvath David J, Kuban Barry D, Sale Shiva, Fukamachi Kiyotaka, Karimov Jamshid H
Department of Biomedical Engineering, Lerner Research Institute, Cleveland Clinic, 9500 Euclid Avenue, Cleveland, OH, 44195, USA.
R1 Engineering LLC, Euclid, OH, USA.
J Artif Organs. 2019 Sep;22(3):256-259. doi: 10.1007/s10047-019-01105-7. Epub 2019 May 10.
This study aimed to evaluate the effects of posture (sitting [lying down]/standing) on hemodynamic and pump-related parameters in calves implanted with our institution's continuous-flow total artificial heart (CFTAH). These parameters were analyzed with posture information in four calves that had achieved the intended 14-, 30-, or 90-day durations of implantation. In each animal, postoperative hourly data gathered throughout the study were used to compare average values with the animal sitting vs. standing. Pump flow became significantly higher in the standing than sitting position at the same pump speed (standing 7.9 ± 0.8, sitting 7.4 ± 1.0 L/min, p = 0.028). Systemic vascular resistance (SVR) and aortic pressure (AoP) were significantly lower in the standing than sitting position (SVR standing 779 ± 145, sitting 929 ± 206 dyne s/cm, p = 0.027; AoP standing 93 ± 7, sitting 103 ± 7 mm Hg, p < 0.001). No substantial change occurred in pulmonary vascular resistance (PVR) or pulmonary arterial pressure (PAP) with posture (PVR standing 161 ± 39, sitting 164 ± 48 dyne s/cm, p = 0.639; PAP standing 32 ± 3, sitting 33 ± 4 mm Hg, p = 0.340). Posture affected some hemodynamic and pump-related parameters in calves with CFTAH, with implications for patients with implanted pumps.
本研究旨在评估姿势(坐[躺]位/站立位)对植入本机构连续流全人工心脏(CFTAH)的小牛的血流动力学和泵相关参数的影响。对4头已达到预期14天、30天或90天植入时长的小牛,结合姿势信息对这些参数进行了分析。在每只动物中,整个研究过程中收集的术后每小时数据用于比较动物站立位与坐位时的平均值。在相同泵速下,站立位时的泵流量显著高于坐位(站立位7.9±0.8,坐位7.4±1.0 L/分钟,p = 0.028)。站立位时的全身血管阻力(SVR)和主动脉压(AoP)显著低于坐位(SVR站立位779±145,坐位929±206达因·秒/厘米,p = 0.027;AoP站立位93±7,坐位103±7毫米汞柱,p < 0.001)。姿势改变时,肺血管阻力(PVR)或肺动脉压(PAP)无显著变化(PVR站立位161±39,坐位164±48达因·秒/厘米,p = 0.639;PAP站立位32±3,坐位33±4毫米汞柱,p = 0.340)。姿势会影响植入CFTAH的小牛的一些血流动力学和泵相关参数,这对植入泵的患者具有启示意义。