经 flash-replenishment 超声造影成像暴露后的啮齿动物肾脏的组织学和血液化学检查。

Histological and blood chemistry examination of the rodent kidney after exposure to flash-replenishment ultrasound contrast imaging.

机构信息

Joint Department of Biomedical Engineering, University of North Carolina Chapel Hill and North Carolina State University, Chapel Hill, NC, USA.

Department of Veterinary Biosciences, The Ohio State University, Columbus, OH, USA.

出版信息

Ultrasonics. 2019 Sep;98:1-6. doi: 10.1016/j.ultras.2019.05.003. Epub 2019 May 10.

Abstract

The purpose of this work is to investigate whether imaging sequences of flash-replenishment contrast enhanced ultrasound (CEUS) of the kidney result in chronic or acute bioeffects. Kidneys of female Fischer 344 rats were imaged using the flash-replenishment technique. Animals were separated into four groups (N = 31). Imaging was conducted with a 4C1 probe, driven by an Acuson Sequoia system with Definity microbubbles as the ultrasound contrast agent. During the flash phase of the imaging sequence, one kidney in each animal was exposed to either a mechanical index (MI) of 1.0 or 1.9. For each MI, half of the animals were sacrificed shortly after imaging (4 h) or after 2 weeks. A blinded veterinary nephropathologist reviewed the histopathology of both the imaged and control (non-imaged) kidney. Blood urea nitrogen (BUN) was measured for each animal prior to imaging and at the time of necropsy. Histopathology assessments in both the 1.0 and 1.9 MI groups revealed no signs of hemorrhage at either the 4-h or 2-week time point. BUN showed minor but statistically significant elevations in both the 1.0 and 1.9 MI groups, but no significant difference was present at the 2-week time point in the 1.0 MI group. All BUN levels (at both time points) remained in the normal range. In conclusion, CEUS with flash-replenishment imaging sequences did not result in kidney bioeffects observable with histology at early or late time points. Increases in BUN levels were observed after imaging, but were minimized when using a moderate MI (1.0).

摘要

本研究旨在探讨肾脏 Flash 补充对比增强超声(CEUS)成像序列是否会导致慢性或急性生物效应。使用 Flash 补充技术对雌性 Fischer 344 大鼠的肾脏进行成像。将动物分为四组(N=31)。使用 4C1 探头进行成像,由 Acuson Sequoia 系统驱动,使用 Definity 微泡作为超声对比剂。在成像序列的 Flash 阶段,每个动物的一侧肾脏暴露于机械指数(MI)为 1.0 或 1.9。对于每个 MI,一半的动物在成像后不久(4 小时)或 2 周后处死。一名盲法兽医肾病学家评估了成像和对照(未成像)肾脏的组织病理学。在成像前和解剖时测量每个动物的血尿素氮(BUN)。在 1.0 和 1.9 MI 组的组织病理学评估中,在 4 小时或 2 周时均未发现任何出血迹象。1.0 和 1.9 MI 组的 BUN 均显示出轻微但具有统计学意义的升高,但在 1.0 MI 组的 2 周时无显著差异。所有 BUN 水平(在两个时间点)均在正常范围内。总之,Flash 补充成像序列的 CEUS 并未在早期或晚期时间点通过组织学观察到肾脏生物效应。在成像后观察到 BUN 水平升高,但在使用中度 MI(1.0)时最小化。

相似文献

2
5
Contrast-assisted destruction-replenishment ultrasound for the assessment of tumor microvasculature in a rat model.
Technol Cancer Res Treat. 2002 Dec;1(6):459-70. doi: 10.1177/153303460200100606.
6
Bioeffects of myocardial contrast microbubble destruction by echocardiography.
Echocardiography. 2002 Aug;19(6):495-500. doi: 10.1046/j.1540-8175.2002.00495.x.
7
Renal retention of lipid microbubbles: a potential mechanism for flank discomfort during ultrasound contrast administration.
J Am Soc Echocardiogr. 2013 Dec;26(12):1474-81. doi: 10.1016/j.echo.2013.08.004. Epub 2013 Sep 12.
8
Real-time contrast ultrasound muscle perfusion imaging with intermediate-power imaging coupled with acoustically durable microbubbles.
J Am Soc Echocardiogr. 2015 Jun;28(6):718-26.e2. doi: 10.1016/j.echo.2015.02.002. Epub 2015 Mar 11.
9
Pulse sequences for uniform perfluorocarbon droplet vaporization and ultrasound imaging.
Ultrasonics. 2014 Sep;54(7):2024-33. doi: 10.1016/j.ultras.2014.05.013. Epub 2014 Jun 6.
10
Influence of Phase Change Droplet Activation and Microbubble Cavitation on the Microenvironment of Hepatocellular Carcinoma.
Ultrasound Med Biol. 2024 Sep;50(9):1387-1394. doi: 10.1016/j.ultrasmedbio.2024.05.015. Epub 2024 Jun 13.

引用本文的文献

本文引用的文献

1
The Dependence of Glomerular Capillary Hemorrhage Induced by Contrast Enhanced Diagnostic Ultrasound on Microbubble Diameter.
Ultrasound Med Biol. 2018 Mar;44(3):613-621. doi: 10.1016/j.ultrasmedbio.2017.11.007. Epub 2017 Dec 26.
2
A Pilot Clinical Study in Characterization of Malignant Renal-cell Carcinoma Subtype with Contrast-enhanced Ultrasound.
Ultrason Imaging. 2017 Mar;39(2):126-136. doi: 10.1177/0161734616666383. Epub 2016 Sep 22.
3
The clinical use of contrast-enhanced ultrasound in the kidney.
Ultrasound. 2016 May;24(2):94-103. doi: 10.1177/1742271X15627185. Epub 2016 Jan 19.
4
A Two-Criterion Model for Microvascular Bio-Effects Induced In Vivo by Contrast Microbubbles Exposed to Medical Ultrasound.
Ultrasound Med Biol. 2016 Jun;42(6):1385-98. doi: 10.1016/j.ultrasmedbio.2016.01.023. Epub 2016 Mar 28.
5
Management of Indeterminate Cystic Kidney Lesions: Review of Contrast-enhanced Ultrasound as a Diagnostic Tool.
Urology. 2016 Jan;87:1-10. doi: 10.1016/j.urology.2015.10.002. Epub 2015 Oct 19.
7
Quantitative volumetric perfusion mapping of the microvasculature using contrast ultrasound.
Invest Radiol. 2010 Oct;45(10):669-74. doi: 10.1097/RLI.0b013e3181ef0a78.
8
Contrast enhanced ultrasound imaging.
J Nucl Cardiol. 2010 Jan-Feb;17(1):106-15. doi: 10.1007/s12350-009-9165-y.
10
In vivo gas body efficacy for glomerular capillary hemorrhage induced by diagnostic ultrasound in rats.
IEEE Trans Biomed Eng. 2010 Jan;57(1):167-74. doi: 10.1109/TBME.2009.2030960. Epub 2009 Aug 25.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验