Ultrasonics. 2019 Sep;98:1-6. doi: 10.1016/j.ultras.2019.05.003. Epub 2019 May 10.
The purpose of this work is to investigate whether imaging sequences of flash-replenishment contrast enhanced ultrasound (CEUS) of the kidney result in chronic or acute bioeffects. Kidneys of female Fischer 344 rats were imaged using the flash-replenishment technique. Animals were separated into four groups (N = 31). Imaging was conducted with a 4C1 probe, driven by an Acuson Sequoia system with Definity microbubbles as the ultrasound contrast agent. During the flash phase of the imaging sequence, one kidney in each animal was exposed to either a mechanical index (MI) of 1.0 or 1.9. For each MI, half of the animals were sacrificed shortly after imaging (4 h) or after 2 weeks. A blinded veterinary nephropathologist reviewed the histopathology of both the imaged and control (non-imaged) kidney. Blood urea nitrogen (BUN) was measured for each animal prior to imaging and at the time of necropsy. Histopathology assessments in both the 1.0 and 1.9 MI groups revealed no signs of hemorrhage at either the 4-h or 2-week time point. BUN showed minor but statistically significant elevations in both the 1.0 and 1.9 MI groups, but no significant difference was present at the 2-week time point in the 1.0 MI group. All BUN levels (at both time points) remained in the normal range. In conclusion, CEUS with flash-replenishment imaging sequences did not result in kidney bioeffects observable with histology at early or late time points. Increases in BUN levels were observed after imaging, but were minimized when using a moderate MI (1.0).
本研究旨在探讨肾脏 Flash 补充对比增强超声(CEUS)成像序列是否会导致慢性或急性生物效应。使用 Flash 补充技术对雌性 Fischer 344 大鼠的肾脏进行成像。将动物分为四组(N=31)。使用 4C1 探头进行成像,由 Acuson Sequoia 系统驱动,使用 Definity 微泡作为超声对比剂。在成像序列的 Flash 阶段,每个动物的一侧肾脏暴露于机械指数(MI)为 1.0 或 1.9。对于每个 MI,一半的动物在成像后不久(4 小时)或 2 周后处死。一名盲法兽医肾病学家评估了成像和对照(未成像)肾脏的组织病理学。在成像前和解剖时测量每个动物的血尿素氮(BUN)。在 1.0 和 1.9 MI 组的组织病理学评估中,在 4 小时或 2 周时均未发现任何出血迹象。1.0 和 1.9 MI 组的 BUN 均显示出轻微但具有统计学意义的升高,但在 1.0 MI 组的 2 周时无显著差异。所有 BUN 水平(在两个时间点)均在正常范围内。总之,Flash 补充成像序列的 CEUS 并未在早期或晚期时间点通过组织学观察到肾脏生物效应。在成像后观察到 BUN 水平升高,但在使用中度 MI(1.0)时最小化。