Grip strength (GS) test is an essential aspect of clinical practice with patients with upper extremity injuries. The random error of GS test was hypothesized to be proportional to the level of GS. The purpose of the current study was to estimate a precise range for the measurement error of GS in patients following traumatic injuries in the upper extremity.
Following traumatic injuries in the upper extremity, 109 participants completed GS tests twice one weekend apart. The Bland-Altman plot analysis was adopted to estimate the precise limits of agreement with 95% confidence interval (CI).
The mean of three consecutive trials had a higher intraclass correlation coefficient of 0.974 (95% CI = 0.963, 0.982) than those of one trial and the mean of the first two trials in injured upper extremities. When GS was ≤20 kg, the upper limit of agreement with 95% CI was estimated as (0.41 × average GS + 1.24), while the lower limit was estimated as (- 0.41 × average GS - 0.39). A table of one-to-one matches between averaged GS ≤ 20 kg and transformed ranges of random errors with 95% certainty was created; the standard error of measurement and minimal detectable change with 95% certainty of GS test were 1.8 and 4.9 kg, respectively. When GS was > 20 kg, the width of agreement with 95% CI ranged from - 4.9 to 5.3 kg, and the standard error of measurement and minimal detectable change with 95% certainty were 1.8 and 5.1 kg, respectively.
The one-to-one match table can be considered as a practical tool to judge a change in GS score is real or due to random errors when it is ≤20 kg.
握力(GS)测试是上肢损伤患者临床实践的重要组成部分。假设 GS 测试的随机误差与 GS 水平成正比。本研究的目的是估计上肢创伤后患者 GS 测量误差的精确范围。
上肢创伤后,109 名参与者在一个周末内完成了两次 GS 测试。采用 Bland-Altman 绘图分析来估计具有 95%置信区间(CI)的精确一致性界限。
与单次试验相比,连续三次试验的组内相关系数更高,为 0.974(95%CI=0.963,0.982),而前两次试验在受伤上肢中的组内相关系数为 0.952(95%CI=0.934,0.965)。当 GS≤20kg 时,95%CI 置信区间的一致性上限估计为(0.41×平均 GS+1.24),下限估计为(-0.41×平均 GS-0.39)。创建了一个 GS≤20kg 时平均 GS 和随机误差转换范围的一对一匹配表;GS 测试的测量误差标准误差和 95%确定性最小可检测变化分别为 1.8 和 4.9kg。当 GS>20kg 时,95%CI 置信区间的一致性范围为-4.9 至 5.3kg,测量误差标准误差和 95%确定性最小可检测变化分别为 1.8 和 5.1kg。
当 GS≤20kg 时,一对一匹配表可作为判断 GS 评分变化是真实的还是由于随机误差的实用工具。