在前列腺癌患者中,常见动态对比增强 MRI 参数的 T1 时间校正的依赖关系中的观察者内和观察者间变异性。

Intra- and inter-observer variability in dependence of T1-time correction for common dynamic contrast enhanced MRI parameters in prostate cancer patients.

机构信息

Christian Doppler Laboratory for Medical Radiation Research for Radiation Oncology, Medical University of Vienna, Austria; Department of Radiotherapy, Medical University of Vienna/AKH Vienna, Austria.

Christian Doppler Laboratory for Medical Radiation Research for Radiation Oncology, Medical University of Vienna, Austria; Department of Biomedical Imaging and Image-Guided Therapy, Medical University and General Hospital of Vienna, Austria.

出版信息

Eur J Radiol. 2019 Jul;116:27-33. doi: 10.1016/j.ejrad.2019.04.015. Epub 2019 Apr 23.

Abstract

BACKGROUND

Dynamic contrast enhanced (DCE) MRI parameters are potential biomarkers to characterise tumour vasculature and distinguish it from the non-cancerous blood vessel system within the prostate. However, the inevitable presence of intra- and inter-observer variabilities is challenging in this context. Additionally, pre-contrast T1-time correction is a prerequisite to gain quantitative DCE parameters in the first place. The current study investigated the effect of individualized T1-time correction on intra- and inter-reader variability for quantitative DCE-parameters in prostatic lesions.

METHODS

In this IRB-approved retrospective study, two experienced radiologists assessed DCE parameters using individually measured (A) and fixed (B) T1-times twice with a time difference of three weeks. The dataset consisted of 35 MRI-guided biopsy-proven prostate cancer lesions. Limits of agreement (LoA) and coefficients of variability (CoV) were calculated to assess intra- and inter-reader variabilities of the parameters.

RESULTS

With exception of k, for all DCE parameters both intra- and inter-reader CoV were smaller in B compared to A. Absolute k values were largely insensitive to T1-time correction induced bias. The mean intra-reader CoVs [5%, 95% percentile] (over all four DCE parameters and both readers) were 6.7% [0.5%, 15.1%] in A and 3.9% [0.2%, 11.0%] in B. The inter-reader CoVs were 9.0% [0.6%, 25.8%] (A) and 7.0% [0.3%, 25.4%] (B).

CONCLUSIONS

T1-time correction has a significant influence on the intra- and inter-reader variability. By applying individually measured T1-time correction, both intra- and inter-observer variability were found to increase. Out of all investigated DCE parameters, k is the most robust to this investigated bias.

摘要

背景

动态对比增强(DCE)MRI 参数是一种潜在的生物标志物,可用于表征肿瘤血管,并将其与前列腺内的非癌性血管系统区分开来。然而,在这种情况下,不可避免地存在观察者内和观察者间的变异性。此外,在获得定量 DCE 参数之前,需要进行预对比 T1 时间校正。本研究旨在探讨个体化 T1 时间校正对前列腺病变定量 DCE 参数的观察者内和观察者间变异性的影响。

方法

在这项经过机构审查委员会批准的回顾性研究中,两名经验丰富的放射科医生使用分别测量的(A)和固定的(B)T1 时间两次评估 DCE 参数,两次评估之间的时间间隔为 3 周。数据集包括 35 例 MRI 引导下经活检证实的前列腺癌病变。通过计算界限值和变异系数(CV)来评估参数的观察者内和观察者间变异性。

结果

除 k 外,所有 DCE 参数的观察者内和观察者间 CV 在 B 中均小于 A。绝对 k 值对 T1 时间校正引起的偏差基本不敏感。在所有四个 DCE 参数和两个读者中,观察者内 CV 的平均值[5%,95%分位数](A)为 6.7%[0.5%,15.1%],B 为 3.9%[0.2%,11.0%]。观察者间 CV 为 9.0%[0.6%,25.8%](A)和 7.0%[0.3%,25.4%](B)。

结论

T1 时间校正对观察者内和观察者间的变异性有显著影响。通过应用分别测量的 T1 时间校正,发现观察者内和观察者间的变异性均增加。在所研究的 DCE 参数中,k 对这种研究偏差最为稳健。