Nancarrow P A, Fellows K E, Lock J E
Department of Radiology, Children's Hospital, Boston, Massachusetts.
Cardiovasc Intervent Radiol. 1987;10(4):226-9. doi: 10.1007/BF02593876.
Gianturco stainless steel coils were discharged into clear plastic tubes of varying sizes and subjected to pulsatile flow in order to study their behavior in a controlled, in vitro environment. When coils are too small for a given vessel, they tend to migrate distally or even proximally. Coils that are too large tend not to reform completely, limiting their cross-sectional area and potentially affecting their occlusive potential. Thus, each coil was tested in tubes of varying sizes and evaluated with regard to positional stability and extent of reformation.
将吉安图尔科不锈钢线圈放入不同尺寸的透明塑料管中,并使其经受脉动流,以便在可控的体外环境中研究它们的行为。当线圈对于给定血管来说太小的时候,它们往往会向远端甚至近端迁移。太大的线圈往往不能完全重塑,限制了它们的横截面积,并可能影响其闭塞潜力。因此,每个线圈都在不同尺寸的管中进行测试,并就位置稳定性和重塑程度进行评估。