多中心评估及改善PET/CT注射渗透率的质量改进举措

Quality Improvement Initiatives to Assess and Improve PET/CT Injection Infiltration Rates at Multiple Centers.

作者信息

Wong Terence Z, Benefield Thad, Masters Shane, Kiser Jackson W, Crowley James, Osborne Dustin, Mawlawi Osama, Barnwell James, Gupta Pawan, Mintz Akiva, Ryan Kelley A, Perrin Steven R, Lattanze Ronald K, Townsend David W

机构信息

Duke University, Durham, North Carolina.

University of North Carolina, Chapel Hill, North Carolina.

出版信息

J Nucl Med Technol. 2019 Dec;47(4):326-331. doi: 10.2967/jnmt.119.228098. Epub 2019 Jun 10.

Abstract

PET/CT radiotracer infiltration is not uncommon and is often outside the imaging field of view. Infiltration can negatively affect image quality, image quantification, and patient management. Until recently, there has not been a simple way to routinely practice PET radiopharmaceutical administration quality control and quality assurance. Our objectives were to quantify infiltration rates, determine associative factors for infiltration, and assess whether rates could be reduced at multiple centers and then sustained. A "design, measure, analyze, improve, and control" quality improvement methodology requiring novel technology was used to try to improve PET/CT injection quality. Teams were educated on the importance of quality injections. Baseline infiltration rates were measured, center-specific associative factors were analyzed, team meetings were held, improvement plans were established and executed, and rates remeasured. To ensure that injection-quality gains were retained, real-time feedback and ongoing monitoring were used. Sustainability was assessed. Seven centers and 56 technologists provided data on 5,541 injections. The centers' aggregated baseline infiltration rate was 6.2% (range, 2%-16%). On the basis of their specific associative factors, 4 centers developed improvement plans and reduced their aggregated infiltration rate from 8.9% to 4.6% ( < 0.0001). Ongoing injection monitoring showed sustainability. Significant variation was found in center- and technologist-level infiltration rates ( < 0.0001 and = 0.0020, respectively). A quality improvement approach with new technology can help centers measure infiltration rates, determine associative factors, implement interventions, and improve and sustain injection quality. Because PET/CT images help guide patient management, the monitoring and improvement of radiotracer injection quality are important.

摘要

正电子发射断层显像/计算机断层扫描(PET/CT)放射性示踪剂浸润并不罕见,且常发生在成像视野之外。浸润会对图像质量、图像定量分析及患者管理产生负面影响。直到最近,还没有一种简单的方法可常规用于PET放射性药物给药的质量控制和质量保证。我们的目标是量化浸润率,确定浸润的相关因素,并评估在多个中心能否降低浸润率并维持较低水平。我们采用了一种需要新技术的“设计、测量、分析、改进和控制”质量改进方法,试图提高PET/CT注射质量。对团队进行了高质量注射重要性的培训。测量了基线浸润率,分析了各中心特定的相关因素,召开了团队会议,制定并执行了改进计划,然后重新测量了浸润率。为确保保持注射质量的提升效果,采用了实时反馈和持续监测。评估了可持续性。七个中心和56名技术人员提供了5541次注射的数据。各中心的总体基线浸润率为6.2%(范围为2% - 16%)。基于其特定的相关因素,4个中心制定了改进计划,其总体浸润率从8.9%降至4.6%(P < 0.0001)。持续的注射监测显示具有可持续性。在中心和技术人员层面的浸润率上发现了显著差异(分别为P < 0.0001和P = 0.0020)。采用新技术的质量改进方法可帮助各中心测量浸润率、确定相关因素、实施干预措施,并提高和维持注射质量。由于PET/CT图像有助于指导患者管理,因此监测和改进放射性示踪剂注射质量很重要。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/9f1a/6894099/b8b5e3e5594e/jnmt228098fig1.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验