Laboratory Medicine, College of Medicine, Korea University, Seongbuk-gu, Seoul, Republic of Korea.
BMJ Open. 2019 Aug 5;9(8):e025438. doi: 10.1136/bmjopen-2018-025438.
To compare the diagnostic performance of three rapid antigen detection tests (RADTs) for group A (GAS).
A hospital-based, cross-sectional, retrospective study.
A comparative study of rapid diagnostic tests for GAS using clinical specimens in a single institute.
225 children in the outpatient clinics ofKorea University Guro Hospitalwith suspicious symptoms were subjected to throat swab sampling. A dual-swab applicator was used. Samples were stored at below -70°C in a 10 mL transport tube containing 1 mL liquid Stuart's transport medium.
All tests were performed in the laboratory by trained clinical laboratory scientists. Sensitivity, specificity, accuracy and kappa index of three RADTs were compared with the reference PCR test and culture results.
Of the 225 patients suspected of having GAS, 67 and 90 were positive for GAS in the culture and PCR tests, respectively. Compared with the reference culture, the sensitivity for GAS was 92.5% (CI 83.4 to 97.5), 71.6% (CI 59.3 to 81.9) and 74.63% (CI 62.5 to 84.4) for careUS Strep A Plus, SD Bioline and BD Veritor, respectively, and the specificity was 97.0% (CI 93.1 to 99.0), 94.6% (CI 90.1 to 97.5) and 92.9% (CI 87.8 to 96.2) for careUS Strep A Plus, SD Bioline and BD Veritor, respectively. Compared with the reference GAS real-time PCR, the sensitivity was 73.3% (CI 62.9 to 82.1), 63.3% (CI 52.5 to 73.2) and 67.8% (CI 57.1 to 77.2) for careUS Strep A Plus, SD Bioline and BD Veritor, respectively, and the specificity was 99.3% (CI 95.9 to 99.9), 100.0% (CI 97.3 to 100.0) and 99.3% (CI 95.9 to 99.9) for careUS Strep A Plus, SD Bioline and BD Veritor, respectively.
The careUS Strep A Plus is a useful test that showed highly comparable results with those of the culture test and superior performances among the three RADTs. The use of RADTs should be encouraged to provide acceptable and fast results using simple equipment.
比较三种快速抗原检测试验(RADT)对 A 组链球菌(GAS)的诊断性能。
基于医院的、横断面的、回顾性研究。
在一家机构内对 GAS 的快速诊断检测进行比较的研究。
225 名在韩国大学古罗医院门诊出现可疑症状的儿童接受了咽喉拭子采样。使用双拭子涂抹器。样本储存在 -70°C 以下,在含有 1mL 液体 Stuart 运输介质的 10mL 运输管中。
所有检测均由经过培训的临床实验室科学家在实验室中进行。比较了三种 RADT 的灵敏度、特异性、准确性和kappa 指数与参考 PCR 检测和培养结果的关系。
在 225 名疑似患有 GAS 的患者中,培养和 PCR 检测结果分别有 67 例和 90 例 GAS 阳性。与参考培养相比,careUS Strep A Plus 对 GAS 的灵敏度为 92.5%(83.4%97.5%),SD Bioline 为 71.6%(59.3%81.9%),BD Veritor 为 74.63%(62.5%84.4%),特异性分别为 97.0%(93.1%99.0%)、94.6%(90.1%97.5%)和 92.9%(87.8%96.2%)。与参考 GAS 实时 PCR 相比,careUS Strep A Plus 的灵敏度为 73.3%(62.9%82.1%),SD Bioline 为 63.3%(52.5%73.2%),BD Veritor 为 67.8%(57.1%77.2%),特异性分别为 99.3%(95.9%99.9%)、100.0%(97.3%100.0%)和 99.3%(95.9%99.9%)。
careUS Strep A Plus 是一种有用的检测方法,其结果与培养试验高度可比,并且在三种 RADT 中表现更为出色。应鼓励使用 RADT,以便使用简单的设备提供可接受和快速的结果。