血管内动脉瘤修复术后移植物特异性血流动力学:四种支架移植物系统的血流特征。

Endograft Specific Haemodynamics After Endovascular Aneurysm Repair: Flow Characteristics of Four Stent Graft Systems.

机构信息

Laboratory for Vascular Simulations, Institute of Vascular Diseases, Larissa, Greece.

Laboratory for Vascular Simulations, Institute of Vascular Diseases, Larissa, Greece; Department of Mathematics, University of Ioannina, Ioannina, Greece.

出版信息

Eur J Vasc Endovasc Surg. 2019 Oct;58(4):538-547. doi: 10.1016/j.ejvs.2019.04.017. Epub 2019 Aug 17.

Abstract

OBJECTIVES

The implication of haemodynamics in the occurrence of complications after endovascular aneurysm repair (EVAR) has been raised in the literature. Different aortic stent graft configurations may lead to different haemodynamic properties. The current study deals with the post-operative haemodynamic variability between four stent graft systems with different structure, material, and type of fixation.

METHODS

Computed tomography data of 32 patients were used, equally distributed among the four endograft groups, namely the AFX, Endurant, Excluder, and Nellix. Velocity, wall shear stress (WSS), and helicity statistics were calculated, in regions around the flow division where disturbances are expected. The haemodynamic data were compared between and within the groups.

RESULTS

The morphology of AAAs pre-operatively did not vary significantly among the four groups. Before the flow division, lowest velocity was observed in Endurant cases and highest in Nellix cases. Endurant induced the lowest peak WSS and Nellix the highest (p = .03). The helicity levels were low in AFX and Nellix cases and high in Endurant and Excluder cases. After the flow division, the trend in the results was preserved. Nellix induced the highest velocity and WSS, followed closely by Excluder and AFX. There was a significant increase of helicity before and after flow division in AFX (p <0.001, R = 0.09) and Nellix (p <0.001) cases.

CONCLUSIONS

It has been shown that different types of endografts induce variable haemodynamic conditions around the flow division. The parallel limb structure, featured by Nellix, seems to induce favourable flow conditions in terms of velocity and WSS, while helical flow before the flow division is suppressed. High WSS is generally considered to be a desirable flow characteristic in endovascular devices, whereas helicity extremes (very low or high) are potentially a negative sign. Endurant, with the stiffer material and the short neck structure, was associated with the lowest blood velocity and WSS values but preserved high helicity levels. The AFX and Excluder, which include the same material, induced similar haemodynamic conditions.

摘要

目的

文献中提出,血管内动脉瘤修复(EVAR)后并发症的发生与血流动力学有关。不同的主动脉支架移植物构型可能导致不同的血流动力学特性。本研究涉及四种不同结构、材料和固定类型的支架移植物系统之间术后血流动力学的变异性。

方法

使用 32 名患者的计算机断层扫描数据,平均分布在四个支架移植物组中,即 AFX、Endurant、Excluder 和 Nellix。在预期发生干扰的分流区域周围计算速度、壁面切应力(WSS)和螺旋度统计数据。在组间和组内比较血流动力学数据。

结果

术前 AAA 的形态在四组之间没有显著差异。在分流前,Endurant 病例的速度最低,Nellix 病例的速度最高。Endurant 引起的峰值 WSS 最低,而 Nellix 引起的峰值 WSS 最高(p=0.03)。AFX 和 Nellix 病例的螺旋度水平较低,Endurant 和 Excluder 病例的螺旋度水平较高。分流后,结果的趋势保持不变。Nellix 引起的速度和 WSS 最高,紧随其后的是 Excluder 和 AFX。AFX(p<0.001,R=0.09)和 Nellix(p<0.001)病例在分流前后的螺旋度显著增加。

结论

不同类型的支架移植物在分流周围诱导不同的血流动力学条件。Nellix 特有的平行支结构似乎在速度和 WSS 方面诱导出有利的流动条件,而分流前的螺旋流则被抑制。高 WSS 通常被认为是血管内装置中理想的流动特征,而螺旋度极值(非常低或非常高)可能是一个负面信号。Endurant 具有更硬的材料和短颈结构,与最低的血流速度和 WSS 值相关,但保持了较高的螺旋度水平。包含相同材料的 AFX 和 Excluder 诱导出相似的血流动力学条件。