应用加速 3D 非对比黑血心血管磁共振技术与压缩感知(CS-DANTE-SPACE)对腹主动脉瘤进行监测。

Surveillance of abdominal aortic aneurysm using accelerated 3D non-contrast black-blood cardiovascular magnetic resonance with compressed sensing (CS-DANTE-SPACE).

机构信息

Department of Radiology and Biomedical Imaging, UCSF, 4150 Clement Street, San Francisco, CA, 94121, USA.

Department of Radiology, Xuanwu Hospital, Beijing, China.

出版信息

J Cardiovasc Magn Reson. 2019 Oct 28;21(1):66. doi: 10.1186/s12968-019-0571-2.

Abstract

BACKGROUND

3D non-contrast high-resolution black-blood cardiovascular magnetic resonance (CMR) (DANTE-SPACE) has been used for surveillance of abdominal aortic aneurysm (AAA) and validated against computed tomography (CT) angiography. However, it requires a long scan time of more than 7 min. We sought to develop an accelerated sequence applying compressed sensing (CS-DANTE-SPACE) and validate it in AAA patients undergoing surveillance.

METHODS

Thirty-eight AAA patients (all males, 73 ± 6 years) under clinical surveillance were recruited for this study. All patients were scanned with DANTE-SPACE (scan time 7:10 min) and CS-DANTE-SPACE (scan time 4:12 min, a reduction of 41.4%). Nine 9 patients were scanned more than 2 times. In total, 50 pairs of images were available for comparison. Two radiologists independently evaluated the image quality on a 1-4 scale, and measured the maximal diameter of AAA, the intra-luminal thrombus (ILT) and lumen area, ILT-to-muscle signal intensity ratio, and the ILT-to-lumen contrast ratio. The sharpness of the aneurysm inner/outer boundaries was quantified.

RESULTS

CS-DANTE-SPACE achieved comparable image quality compared with DANTE-SPACE (3.15 ± 0.67 vs. 3.03 ± 0.64, p = 0.06). There was excellent agreement between results from the two sequences for diameter/area and ILT ratio measurements (ICCs> 0.85), and for quantifying growth rate (3.3 ± 3.1 vs. 3.3 ± 3.4 mm/year, ICC = 0.95.) CS-DANTE-SPACE showed a higher ILT-to-lumen contrast ratio (p = 0.01) and higher sharpness than DANTE-SPACE (p = 0.002). Both sequences had excellent inter-reader reproducibility for quantitative measurements (ICC > 0.88).

CONCLUSION

CS-DANTE-SPACE can reduce scan time while maintaining image quality for AAA imaging. It is a promising tool for the surveillance of patients with AAA disease in the clinical setting.

摘要

背景

3D 非对比高分辨率黑血心血管磁共振(CMR)(DANTE-SPACE)已用于监测腹主动脉瘤(AAA),并通过计算机断层扫描(CT)血管造影进行验证。然而,它需要超过 7 分钟的长扫描时间。我们旨在开发一种应用压缩感知(CS-DANTE-SPACE)的加速序列,并在接受监测的 AAA 患者中进行验证。

方法

本研究招募了 38 名接受临床监测的 AAA 患者(均为男性,73±6 岁)。所有患者均接受 DANTE-SPACE(扫描时间 7:10 分钟)和 CS-DANTE-SPACE(扫描时间 4:12 分钟,减少 41.4%)扫描。9 名患者进行了两次以上扫描。总共获得了 50 对图像进行比较。两名放射科医生独立对图像质量进行 1-4 分的评估,并测量 AAA 的最大直径、腔内血栓(ILT)和管腔面积、ILT-肌肉信号强度比以及 ILT-管腔对比度比。还量化了动脉瘤内/外边界的锐利度。

结果

CS-DANTE-SPACE 实现了与 DANTE-SPACE 相当的图像质量(3.15±0.67 与 3.03±0.64,p=0.06)。两种序列在直径/面积和 ILT 比值测量方面具有极好的一致性(ICC>0.85),并且在量化增长率方面也具有极好的一致性(3.3±3.1 与 3.3±3.4 毫米/年,ICC=0.95)。CS-DANTE-SPACE 显示出更高的 ILT-管腔对比度比(p=0.01)和更高的锐利度(p=0.002)。两种序列的定量测量均具有极好的读者间可重复性(ICC>0.88)。

结论

CS-DANTE-SPACE 可以在保持图像质量的同时减少 AAA 成像的扫描时间。它是一种有前途的工具,可用于临床环境中 AAA 疾病患者的监测。