Suppr超能文献

一种用于经导管封堵房间隔缺损患者器械尺寸选择的新型三维超声心动图方法。

A novel three-dimensional echocardiographic method for device size selection in patients undergoing ASD trans-catheter closure.

作者信息

Roushdy Alaa, El Sayegh Aya, Ali Yasmin Abdelrazek, Attia Hebattalla, El Fiky Azza, El Sayed Maiy

机构信息

Cardiology Department, Congenital and structural heart diseases unit, Ain Shams University hospitals, Cairo, Egypt.

出版信息

Egypt Heart J. 2019 Dec 31;72(1):1. doi: 10.1186/s43044-019-0038-7.

Abstract

BACKGROUND

Proper device size selection is a crucial step for successful ASD device closure. The current gold standard for device size selection is balloon sizing. Balloon sizing can be tedious, time consuming and increase fluoroscopy and procedure times as well as risk of complications. We aimed to establish a simple and accurate method for device size selection using three-dimensional echocardiographic interrogation of the ASD.This is a prospective observational study conducted over a period of 12 months. All patients underwent 2D TTE, three-dimensional echocardiographic assessment of the IAS and transesophageal echocardiogram. Comparison between echocardiographic variables was done using independent sample t test. Linear correlation was established between three dimensional echocardiographic variables and respective variables of device size and 2D TTE and TEE measurements.

RESULTS

The study included 50 patients who underwent successful ASD device closure with properly sized device. There was no significant difference between 3D ASD maximum diameter and all diameters measured by TTE and TEE. There was a strong positive correlation between device size used for closure and both 3D measured ASD area (r = 0.907, P<0.0001) and 3D measured ASD circumference (r = 0.917, P<0.0001). Two regression equations were generated to determine proper device size where Device size = 10.8 + [3.95 x 3D ASD area] and Device size = [3.85 x 3D ASD circumference] -1.02 CONCLUSION: Three-dimensional echocardiogram can provide a simple and accurate method for device size selection in patients undergoing ASD device closure using either 3D derived ASD area or ASD circumference.

摘要

背景

选择合适的装置尺寸是房间隔缺损(ASD)装置封堵成功的关键步骤。目前选择装置尺寸的金标准是球囊扩张测量。球囊扩张测量可能繁琐、耗时,会增加透视时间和手术时间以及并发症风险。我们旨在建立一种使用ASD的三维超声心动图检查来选择装置尺寸的简单而准确的方法。

这是一项为期12个月的前瞻性观察性研究。所有患者均接受二维经胸超声心动图(2D TTE)、房间隔的三维超声心动图评估和经食管超声心动图检查。使用独立样本t检验对超声心动图变量进行比较。在三维超声心动图变量与装置尺寸的相应变量以及2D TTE和TEE测量值之间建立线性相关性。

结果

该研究纳入了50例使用尺寸合适的装置成功进行ASD装置封堵的患者。3D ASD最大直径与TTE和TEE测量的所有直径之间无显著差异。用于封堵的装置尺寸与3D测量的ASD面积(r = 0.907,P<0.0001)和3D测量的ASD周长(r = 0.917,P<0.0001)均呈强正相关。生成了两个回归方程来确定合适的装置尺寸,即装置尺寸 = 10.8 + [3.95×3D ASD面积],装置尺寸 = [3.85×3D ASD周长] - 1.02

结论

三维超声心动图可为使用3D得出的ASD面积或ASD周长进行ASD装置封堵的患者提供一种简单而准确的装置尺寸选择方法。