体外循环中搏动与非搏动血流:溶血程度及临床意义。

Pulsatile Versus Nonpulsatile Flow During Cardiopulmonary Bypass: Extent of Hemolysis and Clinical Significance.

机构信息

From the Department of Anaesthetics and Intensive Care, Singapore General Hospital, Singapore.

Department of Haematology, Royal Papworth Hospital, Cambridge, United Kingdom.

出版信息

ASAIO J. 2020 Sep/Oct;66(9):1025-1030. doi: 10.1097/MAT.0000000000001154.

Abstract

Pulsatile flow has been used during cardiopulmonary bypass (CPB) for decades and its use is increasing with advancing extracorporeal technology. Pulsatile flow generates higher circuit pressures and shear forces than nonpulsatile flow at comparable pump flow and patient mean arterial pressure. Very little is known about the effect this has on erythrocytes. We included 62 adult patients (32 in the pulsatile group and 30 in the nonpulsatile group) undergoing elective coronary artery bypass grafting in this prospective observational study. Blood samples were collected at routine sampling times throughout surgery and were analyzed for the presence of free heme and globin using mass spectroscopy. Patient characteristics, CPB, and aortic cross-clamp times, pump flow as well as patient mean arterial pressure were similar in both groups. Maximum circuit pressure in the pulsatile flow group was statistically significantly higher than that in the nonpulsatile flow group (257.12 vs. 190.64 mmHg, p < 0.0001). Both heme and globin levels were higher in the pulsatile flow group. This reached statistical significance with globin at 30 minutes of CPB and with heme after aortic unclamping. We conclude that pulsatile CPB using roller pumps results in a greater extent of hemolysis. The clinical significance, however, is not yet known.

摘要

搏动性血流在体外循环(CPB)中已经使用了几十年,随着体外技术的不断进步,其使用也在增加。与可比的泵流量和患者平均动脉压相比,搏动性血流在更高的回路压力和剪切力下产生。关于这对红细胞的影响,我们知之甚少。我们在这项前瞻性观察研究中纳入了 62 名接受择期冠状动脉旁路移植术的成年患者(搏动组 32 例,非搏动组 30 例)。在整个手术过程中,按常规时间采集血样,并使用质谱法分析游离血红素和球蛋白的存在情况。两组患者的特征、CPB 和主动脉阻断时间、泵流量以及患者平均动脉压相似。搏动组的最大回路压力明显高于非搏动组(257.12 与 190.64mmHg,p < 0.0001)。搏动组的血红素和球蛋白水平均升高。CPB 30 分钟时球蛋白达到统计学意义,主动脉开放后血红素达到统计学意义。我们得出结论,使用滚压泵的搏动性 CPB 会导致更严重的溶血。然而,其临床意义尚不清楚。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验