Department of Ophthalmology, Kyorin Eye Center, Kyorin University School of Medicine, Mitaka-City, Japan .
Retina. 2021 Feb 1;41(2):445-453. doi: 10.1097/IAE.0000000000002811.
To compare the flow dynamics of beveled-tip to that of flat-tip vitreous cutters.
The aspiration rates of balanced salt solution and swine vitreous were measured for 25-gauge and 27-gauge beveled-tip and flat-tip vitreous cutters. Fluorescent polymer microspheres were mixed with balanced salt solution to make the flow visible. The flow dynamics at maximal cutting rates and reflux of balanced salt solution were measured with video images of a high-speed camera.
The aspiration rates of balanced salt solution of the 25-gauge and 27-gauge beveled-tip cutters were significantly higher than that of the flat-tip cutters at 7,500 cut/min (P = 0.0001, P = 0.038, respectively). The aspiration rate of swine vitreous by the 25-gauge beveled-tip cutter was significantly higher than that of the flat-tip cutters (P = 0.006). The mean aspiration flow in front of the cutter was significantly faster for both beveled-tip than flat-tip cutters (P = 0.003, P = 0.023). The angle of the mean aspiration flow of both beveled-tip cutters was turned to the proximal side (P = 0.004, P = 0.003). The mean reflux flow of both beveled-tip cutters was significantly faster than that of the flat-tip cutters (P = 0.006, P = 0.006).
The beveled-tip cutters have a greater velocity of aspirating frontal and proximal flow resulting in higher aspiration rates and greater reflux flow velocity.
比较斜口和直口玻璃体切割器的流动力学特性。
测量 25G 和 27G 斜口和直口玻璃体切割器对平衡盐溶液和猪玻璃体的抽吸率。将荧光聚合物微球混入平衡盐溶液中,使流动可视化。使用高速摄像机的视频图像测量最大切割速度和平衡盐溶液回流时的流动力学。
7500 转/分时,25G 和 27G 斜口切割器的平衡盐溶液抽吸率明显高于直口切割器(P = 0.0001,P = 0.038)。25G 斜口切割器抽吸猪玻璃体的速率明显高于直口切割器(P = 0.006)。切割器前的平均抽吸流速在斜口切割器均明显快于直口切割器(P = 0.003,P = 0.023)。两种斜口切割器的平均抽吸流角度均向近侧旋转(P = 0.004,P = 0.003)。两种斜口切割器的平均回流流速均明显快于直口切割器(P = 0.006,P = 0.006)。
斜口切割器具有更大的前向和近侧抽吸速度,从而产生更高的抽吸率和更大的回流流速。