Cascade 装置 - 评估线圈向母血管突出的体外测试。

The Cascade device - In vitro tests to assess coil protrusion into the parent vessel.

机构信息

Department of Interventional Neuroradiology, The Royal London Hospital, London, UK.

出版信息

Interv Neuroradiol. 2020 Aug;26(4):494-500. doi: 10.1177/1591019920925708. Epub 2020 May 11.

Abstract

BACKGROUND

Balloon and stent-assisted coiling of aneurysms have increased the number of aneurysms available for endovascular treatment. Newer devices that allow flow within the parent vessel but coverage at the neck have recently entered the market. The Cascade is a new non-occlusive fully retrievable neck-bridging support device that has been designed to provide temporary support during coil embolisation of intracranial aneurysms.

METHODS

Using a silicone aneurysm model three different aneurysms were catheterised with the coiling microcatheter placed in three different positions within each aneurysm - at the neck, centrally, and looped within the aneurysm. Multiple different coils were then deployed within each aneurysm with the Cascade device deployed across the neck to provide protection. In total 480 attempted coilings were performed. Aneurysm flow was used to calculate the change in intra-aneurysmal flow with the Cascade device deployed across the neck of the aneurysm.

RESULTS

We did not observe a single episode of coil protrusion through the Cascade mesh nor did we observe any coil protrusion into the parent vessel when the Cascade was deployed across the neck. There was an average flow velocity reduction of 23% with the Cascade device deployed across the neck of the aneurysm.

CONCLUSION

The Cascade device offers robust protection of the aneurysmal neck and parent vessel as well as inducing significant intra-aneurysmal flow velocity reduction.

摘要

背景

球囊和支架辅助的动脉瘤栓塞术增加了可进行血管内治疗的动脉瘤数量。最近,一些新的装置进入市场,这些装置允许在母血管内流动,同时覆盖在颈部。Cascade 是一种新型的非闭塞性、完全可回收的颈部桥接支撑装置,旨在为颅内动脉瘤的线圈栓塞提供临时支撑。

方法

使用硅胶动脉瘤模型,将三个不同的动脉瘤用微导管进行导管插入,将微导管放置在每个动脉瘤的三个不同位置 - 在颈部、中央和动脉瘤内的环中。然后在每个动脉瘤内部署多个不同的线圈,并在颈部部署 Cascade 装置以提供保护。总共进行了 480 次尝试的线圈栓塞。利用动脉瘤内的血流来计算 Cascade 装置在颈部部署后动脉瘤内的血流变化。

结果

我们没有观察到单个线圈通过 Cascade 网突出的情况,也没有观察到 Cascade 装置在颈部部署时线圈突出到母血管的情况。在颈部部署 Cascade 装置后,平均流速降低了 23%。

结论

Cascade 装置为动脉瘤颈部和母血管提供了强大的保护,同时还能显著降低动脉瘤内的血流速度。

相似文献

1
The Cascade device - In vitro tests to assess coil protrusion into the parent vessel.
Interv Neuroradiol. 2020 Aug;26(4):494-500. doi: 10.1177/1591019920925708. Epub 2020 May 11.
2
Balloon-in-stent technique for the constructive endovascular treatment of "ultra-wide necked" circumferential aneurysms.
Neurosurgery. 2005 Dec;57(6):1218-27; discussion 1218-27. doi: 10.1227/01.neu.0000186036.35823.10.
3
Acutely ruptured basilar artery bifurcation aneurysm, treated with simultaneous Cascade and Comaneci temporary-assisted coiling.
J Neurointerv Surg. 2021 Feb;13(2):196. doi: 10.1136/neurintsurg-2020-016320. Epub 2020 Jul 21.
4
Early clinical experience with Cascade: a novel temporary neck bridging device for embolization of intracranial aneurysms.
J Neurointerv Surg. 2020 Mar;12(3):303-307. doi: 10.1136/neurintsurg-2019-015338. Epub 2019 Sep 21.
5
Novel non-occlusive temporary endoluminal neck protection device to assist in the treatment of aneurysms in a canine model.
J Neurointerv Surg. 2011 Sep;3(3):214-8. doi: 10.1136/jnis.2010.003913. Epub 2011 Jan 25.
7
Stents for intracranial wide-necked aneurysms: more than mechanical protection.
Neuroradiology. 2008 Dec;50(12):991-8. doi: 10.1007/s00234-008-0460-0. Epub 2008 Sep 20.
8
In vitro studies for stent-assisted coiling of terminus aneurysms by straight-on intra-aneurysmal stent deployment.
World Neurosurg. 2013 Jul-Aug;80(1-2):120-6. doi: 10.1016/j.wneu.2013.01.014. Epub 2013 Jan 5.

引用本文的文献

1
The Stream Device-A Retrospective Review of 51 Cases.
J Clin Med. 2023 Oct 6;12(19):6384. doi: 10.3390/jcm12196384.
2
Application of the Neuroform Atlas Stent in Intracranial Aneurysms: Current Status.
Front Neurol. 2022 Mar 25;13:829143. doi: 10.3389/fneur.2022.829143. eCollection 2022.

本文引用的文献

2
Flow Diversion for Treatment of Acutely Ruptured Intracranial Aneurysms : A Single Center Experience from 45 Consecutive Cases.
Clin Neuroradiol. 2020 Dec;30(4):835-842. doi: 10.1007/s00062-019-00846-5. Epub 2019 Nov 4.
3
Early clinical experience with Cascade: a novel temporary neck bridging device for embolization of intracranial aneurysms.
J Neurointerv Surg. 2020 Mar;12(3):303-307. doi: 10.1136/neurintsurg-2019-015338. Epub 2019 Sep 21.
4
The use of flow diverters to treat small (≤5 mm) ruptured, saccular aneurysms.
Surg Neurol Int. 2018 Oct 30;9:216. doi: 10.4103/sni.sni_243_18. eCollection 2018.
5
Coiling of ruptured, wide-necked basilar tip aneurysm using double Comaneci technique.
BMJ Case Rep. 2018 May 18;2018:bcr-2017-222703. doi: 10.1136/bcr-2017-222703.
6
Stent placement in patients with acute subarachnoid haemorrhage: when is it justified?
Neuroradiology. 2018 Jul;60(7):735-744. doi: 10.1007/s00234-018-2020-6. Epub 2018 Apr 11.
7
Treatment of blood blister aneurysms of the internal carotid artery with flow diversion.
J Neurointerv Surg. 2018 Nov;10(11):1074-1078. doi: 10.1136/neurintsurg-2017-013701. Epub 2018 Feb 24.
8
Early experience with a temporary bridging device (Comaneci) in the endovascular treatment of ruptured wide neck aneurysms.
J Neurointerv Surg. 2018 Oct;10(10):978-982. doi: 10.1136/neurintsurg-2017-013641. Epub 2018 Feb 6.
9
Intra-aneurysmal flow disruption after implantation of the Medina® Embolization Device depends on aneurysm neck coverage.
PLoS One. 2018 Feb 6;13(2):e0191975. doi: 10.1371/journal.pone.0191975. eCollection 2018.
10
Understanding Angiography-Based Aneurysm Flow Fields through Comparison with Computational Fluid Dynamics.
AJNR Am J Neuroradiol. 2017 Jun;38(6):1180-1186. doi: 10.3174/ajnr.A5158. Epub 2017 Apr 6.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验