开发一种用于液相色谱/质谱联用的快速切换双(ESI/APCI)离子源。

Development of a fast-switching dual (ESI/APCI) ionization source for liquid chromatography/mass spectrometry.

机构信息

Applied Analytical Chemistry, University of Duisburg-Essen, Universitaetsstrasse 5, Essen, 45141, Germany.

出版信息

Rapid Commun Mass Spectrom. 2020 Sep 15;34(17):e8845. doi: 10.1002/rcm.8845.

Abstract

RATIONALE

High-throughput liquid chromatography/mass spectrometry (LC/MS) is an increasing topic in analytical chemistry. Especially the idle time of a mass spectrometer should be reduced for an efficient and cost-saving use. Therefore, a fast-switching dual ion source was developed, which uses the most important ionization techniques at atmospheric pressure, electrospray ionization (ESI) and atmospheric pressure chemical ionization (APCI), with one or more LC systems.

METHODS

The performance of the developed ion source is shown by infusion experiments and chromatographic analyses of different standard substances. A high-throughput method is demonstrated by coupling two UHPLC systems to the dual ion source with a triple quadrupole mass spectrometer.

RESULTS

No decrease in the ion abundance and a stable performance of the mass spectrometer are presented while using the dual ion source. Instrumental limits of detection are 30 ng L for testosterone using ESI and 1 μg L for vitamin D using APCI. A fast switching between two UHPLC systems and the dual ion source leads to a high sample throughput of 50 samples in 75 min with relative standard deviations for testosterone and vitamin D of 1.5% and 3.8%, respectively.

CONCLUSIONS

This work presents the development of a dual ESI and APCI ion source operating simultaneously or in switched mode. The results show sensitive and reliable performance as well as the hyphenation to one or more HPLC systems.

摘要

原理

高通量液相色谱/质谱联用(LC/MS)是分析化学中日益热门的话题。特别是为了高效和节约成本,应尽量减少质谱仪的空闲时间。因此,开发了一种快速切换的双离子源,它使用大气压下最重要的两种电离技术,即电喷雾电离(ESI)和大气压化学电离(APCI),与一个或多个 LC 系统联用。

方法

通过对不同标准物质进行注入实验和色谱分析,展示了所开发离子源的性能。通过将两个 UHPLC 系统与双离子源和三重四极杆质谱仪耦合,展示了一种高通量方法。

结果

使用双离子源时,离子丰度没有下降,质谱仪的性能也很稳定。ESI 检测睾酮的仪器检出限为 30ng/L,APCI 检测维生素 D 的仪器检出限为 1μg/L。双离子源和两个 UHPLC 系统之间的快速切换可实现 50 个样品在 75 分钟内的高样品通量,睾酮和维生素 D 的相对标准偏差分别为 1.5%和 3.8%。

结论

本工作介绍了一种同时或切换模式运行的双 ESI 和 APCI 离子源的开发。结果表明,该离子源具有灵敏、可靠的性能,并且可与一个或多个 HPLC 系统联用。