International consensus recommends a set of continuous glucose monitoring (CGM) metrics to assess quality of diabetes therapy. The impact of individual CGM sensors on these metrics has not been thoroughly studied yet. This post hoc analysis aimed at comparing time in specific glucose ranges, coefficient of variation (CV) of glucose concentrations, and glucose management indicator (GMI) between different CGM systems and different sensors of the same system.
A total of 20 subjects each wore two Dexcom G5 (G5) sensors and two FreeStyle Libre (FL) sensors for 14 days in parallel. Times in ranges, GMI, and CV were calculated for each 14-day sensor experiment, with up to four sensor experiments per subject. Pairwise differences between different sensors of the same CGM system as well as between sensors of different CGM system were calculated for these metrics.
Pairwise differences between sensors of the same model showed larger differences and larger variability for FL than for G5, with some subjects showing considerable differences between the two sensors. When pairwise differences between sensors of different CGM models were calculated, substantial differences were found in some subjects (75th percentiles of differences of time spent <70 mg/dL: 5.0%, time spent >180 mg/dL: 9.2%, and GMI: 0.42%).
Relevant differences in CGM metrics between different models of CGM systems, and between different sensors of the same model, worn by the same study subjects were found. Such differences should be taken into consideration when these metrics are used in the treatment of diabetes.
国际共识推荐了一套连续血糖监测(CGM)指标,用于评估糖尿病治疗的质量。但单个 CGM 传感器对这些指标的影响尚未得到充分研究。本研究旨在比较不同 CGM 系统和同一系统的不同传感器之间的特定血糖范围内的时间、血糖浓度变异系数(CV)和血糖管理指标(GMI)。
共有 20 名受试者,每位受试者同时佩戴两个 Dexcom G5(G5)传感器和两个 FreeStyle Libre(FL)传感器,持续 14 天。每个 14 天的传感器实验计算了时间范围、GMI 和 CV,每个受试者最多有四个传感器实验。对同一 CGM 系统的不同传感器以及不同 CGM 系统的传感器之间的这些指标进行了成对差异比较。
同一型号传感器之间的成对差异显示,FL 的差异较大且变异较大,而 G5 的差异较小且变异较小,一些受试者两个传感器之间的差异较大。当计算不同 CGM 模型传感器之间的成对差异时,一些受试者发现差异较大(<70mg/dL 的时间百分比差异的第 75 个百分位数:5.0%,>180mg/dL 的时间百分比差异:9.2%,GMI 差异:0.42%)。
同一研究对象佩戴的不同 CGM 系统模型之间以及同一模型的不同传感器之间的 CGM 指标存在相关差异。在糖尿病治疗中使用这些指标时,应考虑这些差异。