Abstract

OBJECTIVES

The aim of this study was to characterize national variation in radionuclide calibrator activity response to a single National Institute of Standards and Technology (NIST) traceable reference Ge source used as a surrogate for F at clinical PET centres in England using National Physical Laboratory approved techniques.

METHODS

Readings from 20 instruments at 13 centres using local F and Ge factor settings were recorded with the source located in vial and syringe positions. Ten repeat measurements were conducted to investigate repeatability using % coefficient of variability (COV). Comparison ratios to investigate accuracy were made between calibrator responses and decay-corrected NISTref reference activity for syringe and vial position measurements.

RESULTS

The maximum %COV was 0.79%, while 90, 95 and 80% of calibrators conformed to 5% accuracy for F syringe, Ge syringe and Ge vial position readings, respectively. We revealed a trend towards reduced bias in measurements using Veenstra devices for F and using Capintec devices for Ge factor settings.

CONCLUSIONS

This study demonstrated good repeatability in local device measurements. In total, 70% of English calibrators tested and 88% of all measurements performed achieved 5% accuracy. While statistically significant bias was exhibited between different vendor equipment dependent upon radioisotope selected, our study recommends regular traceability checks for optimum instrument performance conducted within National Metrology Institutes guidelines.

摘要

目的

本研究旨在利用国家物理实验室批准的技术,描述在英格兰临床正电子发射断层扫描(PET)中心,放射性核素校准仪对单个可溯源至美国国家标准与技术研究院(NIST)的参考锗源(用作临床PET中心氟的替代物)的活性响应的全国差异。

方法

使用当地氟和锗的因子设置,记录了13个中心20台仪器在源位于小瓶和注射器位置时的读数。进行了十次重复测量,以使用变异系数百分比(COV)研究重复性。在校准仪响应与注射器和小瓶位置测量的经衰变校正的NISTref参考活性之间进行比较比,以研究准确性。

结果

最大COV百分比为0.79%,而分别有90%、95%和80%的校准仪在氟注射器、锗注射器和锗小瓶位置读数方面符合5%的准确度。我们发现,使用Veenstra设备进行氟测量以及使用Capintec设备进行锗因子设置时,测量偏差有减小的趋势。

结论

本研究表明本地设备测量具有良好的重复性。总体而言,测试的英国校准仪中有70%以及所有测量中有88%达到了5%的准确度。虽然根据所选放射性同位素的不同,不同供应商设备之间存在统计学上显著的偏差,但我们的研究建议在国家计量机构的指导方针内定期进行溯源检查,以实现最佳仪器性能。