Jacobs E, Anzalone T, Sarkozi L
Department of Chemistry, Mount Sinai Hospital, New York, NY.
Clin Chem. 1988 Mar;34(3):585-8.
We evaluated the analytical performance of the immunoprecipitation technique for quantification of creatine kinase (CK; EC 2.7.3.2) isoenzyme MB, focusing on specimens with increased blank activities. The samples we studied had blank values that were equal to or greater than either 15 U/L or the uncorrected CK-MB value. Of 134 specimens selected, 16 and 21 contained macro CK type 1 and CK-BB, respectively. All samples containing macro CK type 1 gave negative CK-MB values, even though four contained significant quantities of CK-MB. Correcting the blank value for the fluid displacement of the second antibody suspension eliminated all but three of the negative results. However, 45 specimens, negative for CK-MB by electrophoresis, gave CK-MB values ranging from 1 to 21 U/L. Furthermore, 11 of the specimens containing macro-CK type 1 were in this group. Thus, the presence of macro-CK type 1 definitely interferes with the immunoprecipitation technique. Our results indicate the need to re-evaluate the diagnostic accuracy of the immunoprecipitation technique, especially when the un-adjusted blank activities are increased.
我们评估了免疫沉淀技术在定量肌酸激酶(CK;EC 2.7.3.2)同工酶MB方面的分析性能,重点关注空白活性增加的标本。我们研究的样本空白值等于或大于15 U/L或未校正的CK-MB值。在选取的134个标本中,分别有16个和21个含有1型巨CK和CK-BB。所有含有1型巨CK的样本CK-MB值均为阴性,尽管其中4个含有大量的CK-MB。校正第二抗体悬浮液的液体置换空白值后,除3个阴性结果外,其余均消除。然而,45个经电泳检测CK-MB为阴性的标本,其CK-MB值在1至21 U/L之间。此外,该组中有11个标本含有1型巨CK。因此,1型巨CK的存在确实会干扰免疫沉淀技术。我们的结果表明,需要重新评估免疫沉淀技术的诊断准确性,尤其是在未调整的空白活性增加时。