He Chuan, Hu Simiao, Zhou Wanxing
The first affiliated hospital of Guangdong Pharmaceutical University, Guangdong, Peopleś Republic of China Guangdong Metabolic Disease Research Center of Integrated Chinese and Western Medicine, Guangdong, Peopleś Republic of China.
PeerJ. 2020 Sep 15;8:e9941. doi: 10.7717/peerj.9941. eCollection 2020.
This study aimed to develop an analytical method using liquid chromatography tandem mass spectrometry (LC-MS/MS) for the determination of angiotensin (Ang) I, Ang (1-9), Ang II, Ang (1-7), Ang (1-5), Ang III, Ang IV in human umbilical vein endothelial cell (HUVEC) culture supernatant.
HUVEC culture supernatant was added with gradient concentrations (0.05-1,000 ng/ml) of standard solutions of the Ang peptides. These samples underwent C18 solid-phase extraction and separation using a preconcentration nano-liquid chromatography mass spectrometry system. The target peptides were detected by a Q Exactive quadrupole orbitrap high-resolution mass spectrometer in the parallel reaction monitoring mode. Ang converting enzyme (ACE) in HUVECs was silenced to examine Ang I metabolism.
The limit of detection was 0.1 pg for Ang II and Ang III, and 0.5 pg for Ang (1-9), Ang (1-7), and Ang (1-5). The linear detection range was 0.1-2,000 pg (0.05-1,000 ng/ml) for Ang II and Ang III, and 0.5-2,000 pg (0.25-1,000 ng/ml) for Ang (1-9) and Ang (1-5). Intra-day and inter-day precisions (relative standard deviation) were <10%. Ang II, Ang III, Ang IV, and Ang (1-5) were positively correlated with ACE expression by HUVECs, while Ang I, Ang (1-7), and Ang (1-9) were negatively correlated.
The nanoflow liquid chromatography-parallel reaction monitoring mass spectrometry-based methodology established in this study can evaluate the Ang peptides simultaneously in HUVEC culture supernatant.
本研究旨在开发一种采用液相色谱串联质谱法(LC-MS/MS)测定人脐静脉内皮细胞(HUVEC)培养上清液中血管紧张素(Ang)I、Ang(1-9)、Ang II、Ang(1-7)、Ang(1-5)、Ang III、Ang IV的分析方法。
向HUVEC培养上清液中加入梯度浓度(0.05-1000 ng/ml)的Ang肽标准溶液。这些样品采用C18固相萃取,并使用预浓缩纳升液相色谱质谱系统进行分离。目标肽通过Q Exactive四极杆轨道阱高分辨率质谱仪在平行反应监测模式下进行检测。沉默HUVEC中的血管紧张素转换酶(ACE)以研究Ang I的代谢。
Ang II和Ang III的检测限为0.1 pg,Ang(1-9)、Ang(1-7)和Ang(1-5)的检测限为0.5 pg。Ang II和Ang III的线性检测范围为0.1-2000 pg(0.05-1000 ng/ml),Ang(1-9)和Ang(1-5)的线性检测范围为0.5-2000 pg(0.25-1000 ng/ml)。日内和日间精密度(相对标准偏差)<10%。Ang II、Ang III、Ang IV和Ang(1-5)与HUVECs中ACE的表达呈正相关,而Ang I、Ang(1-7)和Ang(1-9)与ACE的表达呈负相关。
本研究建立的基于纳升液相色谱-平行反应监测质谱法的方法可同时评估HUVEC培养上清液中的Ang肽。