Evaluation of the Alinity m HIV-1 assay for the quantification of HIV-1 RNA plasma viral load in a high-throughput molecular laboratory in South Africa.

机构信息

Lancet Laboratories, Johannesburg, South Africa.

Lancet Laboratories, Johannesburg, South Africa.

出版信息

J Clin Virol. 2020 Nov;132:104644. doi: 10.1016/j.jcv.2020.104644. Epub 2020 Sep 18.

Abstract

BACKGROUND

Regular HIV-1 viral load monitoring forms an essential part of any successful HIV-1 treatment programme. Abbott Molecular recently released the Alinity m HIV-1 assay to be run on the Alinity m System, a fully automated, continuous and random access analyser using ReadiFlex™ technology.

OBJECTIVES

Our study investigated the performance of the Alinity m HIV-1 assay in comparison to the cobas® HIV-1 test in a high-throughput molecular laboratory.

STUDY DESIGN

We compared the performance of the Alinity m HIV-1 assay with the cobas® HIV-1 test, performed on both the cobas® 4800 and cobas® 6800 systems at three clinically relevant thresholds (50, 200 and 1000 cp/mL).

RESULTS

Excellent correlation (r = 0.98) and agreement (mean bias -0.004 Log cp/mL) was achieved between the cobas® 4800 and Alinity m HIV-1 assay. While there was good correlation between the Alinity m HIV-1 assay and the cobas® 6800 (r = 0.99), Bland-Altman analysis indicated that the cobas® 6800 on average measured 0.22 Log cp/mL higher than the Alinity m HIV-1 assay across the dynamic range. Percentage agreement was excellent at the 200 cp/mL and 1000 cp/mL thresholds and was slightly lower at 50 cp/mL in comparison with the cobas® systems.

CONCLUSIONS

The Alinity m HIV-1 assay compared well with the cobas® HIV-1 test on both the cobas® 4800 and cobas® 6800 systems in a high-throughput molecular laboratory in South Africa, a low- to middle-income country.

摘要

背景

定期进行 HIV-1 病毒载量监测是任何成功的 HIV-1 治疗计划的重要组成部分。雅培分子公司最近发布了 Alinity m HIV-1 检测试剂盒,可在 Alinity m 系统上运行,该系统是一种使用 ReadiFlex™ 技术的全自动、连续和随机访问分析仪。

目的

我们的研究比较了高通量分子实验室中 Alinity m HIV-1 检测试剂盒与 cobas® HIV-1 检测试剂盒的性能。

研究设计

我们比较了 Alinity m HIV-1 检测试剂盒与 cobas® HIV-1 检测试剂盒在三个临床相关阈值(50、200 和 1000 cp/mL)上的性能。

结果

Alinity m HIV-1 检测试剂盒与 cobas® 4800 系统的相关性极好(r = 0.98),一致性也很好(平均偏差为-0.004 Log cp/mL)。Alinity m HIV-1 检测试剂盒与 cobas® 6800 系统的相关性也很好(r = 0.99),但 Bland-Altman 分析表明,在整个动态范围内, cobas® 6800 系统的平均测量值比 Alinity m HIV-1 检测试剂盒高 0.22 Log cp/mL。在 200 cp/mL 和 1000 cp/mL 阈值下,百分比一致性非常好,而与 cobas® 系统相比,在 50 cp/mL 时稍低。

结论

在南非的一个高通量分子实验室中, Alinity m HIV-1 检测试剂盒在 cobas® 4800 和 cobas® 6800 系统上与 cobas® HIV-1 检测试剂盒相比表现良好,南非是一个中低收入国家。