Suppr超能文献

通过脉冲多普勒超声测量的血流速度和收缩期时间间期:测量的可重复性

Blood flow velocity and systolic time intervals measured by pulsed Doppler ultrasound: reproducibility of measurements.

作者信息

Lang-Jensen T

机构信息

Department of Anaesthesia, Rigshospitalet, Copenhagen, Denmark.

出版信息

Cardiovasc Res. 1987 Aug;21(8):582-6. doi: 10.1093/cvr/21.8.582.

Abstract

The reproducibility of measurements of blood flow velocity in the aorta and systolic time intervals using pulsed ultrasound Doppler was determined in 10 subjects. Measurements were made in both the ascending aorta and the aortic arch, and instant maximum as well as instant mean velocities were measured. The smallest variation was found when the maximum velocity was measured in the ascending aorta. The coefficient of variation in measurements was between 6.9% and 13.1%, when determined at intervals of 10-15 min. Again the smallest variation was when the maximum velocity was measured in the ascending aorta. The intraobserver variation was smaller than the interobserver variation (6.9% compared with 10.6%) in the ascending aorta. Both the intraobserver and interobserver variations in systolic time intervals were very small. The mean coefficients of variation were 1.7% for the pre-ejection period (PEP), 1.7% for left ventricular ejection time (LVET), 1.4% for electromechanical systole (QS2), and 2.9% for the PEP:LVET ratio. The mean values of pre-ejection period and left ventricular ejection time were 154 ms and 377 ms, which are a little longer and shorter respectively than measurements made by phonocardiography and pulse pressure monitoring, whereas QS2 was the same (531 ms). The interobserver variations were 2.6%, 2.0%, 1.4%, and 3.9% respectively. It is concluded that pulsed Doppler ultrasound is a reliable method of measuring blood flow velocity and systolic time intervals. Changes in blood flow velocity of 10-12% can be detected by this method.

摘要

在10名受试者中测定了使用脉冲超声多普勒测量主动脉血流速度和收缩期时间间期的可重复性。在升主动脉和主动脉弓均进行了测量,并测量了即时最大速度和即时平均速度。在升主动脉测量最大速度时发现变化最小。当每隔10 - 15分钟进行测定时,测量的变异系数在6.9%至13.1%之间。同样,在升主动脉测量最大速度时变异最小。在升主动脉中,观察者内变异小于观察者间变异(分别为6.9%和10.6%)。收缩期时间间期的观察者内和观察者间变异均非常小。射血前期(PEP)的平均变异系数为1.7%,左心室射血时间(LVET)为1.7%,机电收缩期(QS2)为1.4%,PEP:LVET比值为2.9%。射血前期和左心室射血时间的平均值分别为154毫秒和377毫秒,这分别比通过心音图和脉压监测测量的值略长和略短,而QS2相同(531毫秒)。观察者间变异分别为2.6%、2.0%、1.4%和3.9%。结论是脉冲多普勒超声是测量血流速度和收缩期时间间期的可靠方法。该方法可检测到10 - 12%的血流速度变化。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验