Department of Spine Surgery, Beijing Ji Shui Tan Hospital, Beijing, China.
Spine (Phila Pa 1976). 2021 Jul 15;46(14):E760-E768. doi: 10.1097/BRS.0000000000003902.
An experimental investigation of a robot-assisted ultrasonic osteotome applied to vertebral cancellous bone.
The aim of this study was to investigate the effect of various ultrasonic parameter settings on temperature in the drilling site and penetration time and determine the most suitable parameters for efficient and safe robot-based ultrasonically assisted bone drilling in spinal surgery.
A robot-assisted ultrasonic osteotome device may be safe and effective for spinal drilling.
Sixty specimens of bovine vertebral cancellous were randomly assigned to one of six groups, which varied by mode of ultrasonic vibration (L-T and L) and feed rate (one percent [0.8 mm/s], two percent [1.6 mm/s], and three pecent [2.4 mm/s]). Maximum temperature in the drilling site and penetration time was recorded.
Maximum temperature in the drilling site decreased as output power increased for L-T and L modes, was significantly lower for L-T compared to L mode at each feed rate and power setting, was significantly different at feed rates of 1.6 mm/s versus 0.8 mm/s and 2.4 mm/s versus 0.8 mm/s for L-T mode at an output power of 60 W and 84 W, but was not influenced by feed rate for L mode. Penetration time did not significantly improve as output power increased for both L-T and L modes, was significantly decreased with increased feed rates, but was not significantly different between L-T and L modes.
The optimal parameters for applying a robot-assisted ultrasonic osteotome to vertebral cancellous bone are L-T mode, maximum output power of 120 W, and maximum feed rate of 2.4 mm/s.Level of Evidence: 4.
机器人辅助超声骨刀在松质骨椎骨中的应用的实验研究。
本研究旨在探讨不同超声参数设置对钻孔部位温度、穿透时间的影响,并确定最适合机器人辅助超声骨钻脊柱手术中高效、安全的参数。
机器人辅助超声骨刀装置可能是安全有效的脊柱钻孔方法。
将 60 个牛松质骨标本随机分为六组,组间超声振动模式(L-T 和 L)和进给率(1%[0.8mm/s]、2%[1.6mm/s]和 3%[2.4mm/s])不同。记录钻孔部位的最高温度和穿透时间。
L-T 和 L 模式下,输出功率增加时钻孔部位的最高温度降低,L-T 模式在各进给率和功率设置下均显著低于 L 模式,L-T 模式在输出功率为 60 W 和 84 W 时,进给率为 1.6mm/s 与 0.8mm/s 和 2.4mm/s 与 0.8mm/s 之间差异显著,但 L 模式不受进给率影响。L-T 和 L 模式下,随着输出功率的增加,穿透时间均未显著改善,随着进给率的增加而显著缩短,但 L-T 和 L 模式之间无显著差异。
将机器人辅助超声骨刀应用于松质骨椎骨的最佳参数为 L-T 模式、最大输出功率 120 W 和最大进给率 2.4mm/s。
4 级。