Pathology, Department of Medicine & Surgery, Università di Milano-Bicocca, Via Cadore 48, Monza 20900, Italy.
Department of Experimental Oncology, European Institute of Oncology, Via Adamello 16, Milan 20139, Italy.
Biotechniques. 2021 Mar;70(3):137-148. doi: 10.2144/btn-2020-0149. Epub 2021 Feb 5.
Antibody validation for tissue staining is required for reproducibility; criteria to ensure validity have been published recently. The majority of these recommendations imply the use of routinely processed (formalin-fixed, paraffin-embedded) tissue. We applied to lightly fixed frozen sections a panel of 126 antibodies validated for formalin-fixed, paraffin-embedded tissue with extended criteria. Less than 30% of the antibodies performed as expected with all fixations. 35% preferred one fixation over another, 13% gave nonspecific staining and 23% did not stain at all. Individual antibody variability of the paratope's fitness for the fixed antigen may be the cause. Revalidation of established antibody panels is required when they are applied to sections whose fixation and processing are different from the tissue where they were initially validated.
抗体验证对于组织染色的重现性至关重要;最近已经发布了确保有效性的标准。这些建议中的大多数都暗示使用常规处理(福尔马林固定、石蜡包埋)的组织。我们将一套经过扩展标准验证的用于福尔马林固定、石蜡包埋组织的 126 种抗体应用于轻度固定的冷冻切片。不到 30%的抗体在所有固定剂下表现出预期的效果。35%的抗体更倾向于一种固定剂而不是另一种,13%的抗体产生非特异性染色,23%的抗体根本不染色。可能是抗原固定后抗原表位的适应性的抗体变构体的个体抗体的变异性导致了这种情况。当将已建立的抗体试剂盒应用于与最初验证时的组织固定和处理方式不同的切片时,需要对其进行重新验证。