Suppr超能文献

经导管主动脉瓣置换术的连续无创动脉血压监测。

Continuous Noninvasive Arterial Pressure Monitoring for Transcatheter Aortic Valve Replacement.

机构信息

Department of Anesthesiology, Pain Medicine, and Critical Care Medicine, Massachusetts General Hospital, Harvard Medical School, Boston, MA; Department of Critical Care Medicine, Massachusetts General Hospital Heart Center, Massachusetts General Hospital, Harvard Medical School, Boston, MA.

Department of Anesthesiology, Pain Medicine, and Critical Care Medicine, Massachusetts General Hospital, Harvard Medical School, Boston, MA.

出版信息

J Cardiothorac Vasc Anesth. 2021 Jul;35(7):2026-2033. doi: 10.1053/j.jvca.2021.01.012. Epub 2021 Jan 12.

Abstract

OBJECTIVES

The objective of the present study, which was conducted in patients undergoing transcatheter aortic valve replacement, was to compare continuous noninvasive arterial pressure measured with the ClearSight device (Edwards Lifesciences, Irvine, CA) with invasive radial artery pressure used as the reference method. The authors hypothesized that the ClearSight device is an accurate, precise, safe, and efficient method for arterial blood pressure measurement comparable with an invasive radial arterial line.

DESIGN

The study included the retrospective review of 20 consecutive patients scheduled for elective transcatheter aortic valve replacement with the SAPIEN 3 transcatheter heart valve (Edwards Lifesciences) at a single tertiary academic hospital, who underwent monitoring with both the ClearSight device and an invasive radial arterial pressure line. The patients underwent transcatheter aortic valve replacement from October to December 2019.

SETTING

Single tertiary academic medical center.

PARTICIPANTS

The study comprised 20 patients, with 2,243 unique blood pressure data points.

INTERVENTIONS

None.

MEASUREMENTS AND MAIN RESULTS

A statistically significant correlation between the invasive radial arterial pressure line and the ClearSight device was observed for systolic blood pressure (correlation coefficient 0.86; p < 0.001), diastolic blood pressure (correlation coefficient 0.56; p < 0.001), and mean arterial pressure (correlation coefficient 0.78; p < 0.001). Bland-Altman analysis was used to assess the agreement of systolic blood pressure, diastolic blood pressure, and mean arterial pressure between the two methods. Results for systolic blood pressure between the arterial line and ClearSight device were as follows: bias = 9.8 ± 10.1, percentage bias = 7.6%, and mean error = 15.8%. Results for diastolic blood pressure between the arterial line and ClearSight device were as follows: bias = -5.9 ± 7.8, percentage bias = 10.7%, and mean error = 28.4%. Results for mean arterial pressure between the arterial line and ClearSight device were as follows: bias = 0.3 ± 7.4, percentage bias = 0.4%, and mean error = 18.3%. The concordance rates of systolic blood pressure, diastolic blood pressure, and mean arterial pressure were 100%, 95.1%, and 98.8%, respectively.

CONCLUSIONS

The accuracy, agreement, and precision of the ClearSight device were convincing for mean arterial pressure, systolic blood pressure, and diastolic blood pressure for patients with severe aortic stenosis undergoing elective transcatheter aortic valve replacement.

摘要

目的

本研究旨在比较经导管主动脉瓣置换术患者中连续无创动脉压测量的 ClearSight 装置(爱德华兹生命科学公司,加利福尼亚州欧文)与作为参考方法的有创桡动脉压。作者假设 ClearSight 装置是一种准确、精确、安全且有效的动脉血压测量方法,与有创桡动脉线相当。

设计

该研究回顾性分析了 20 例连续接受经导管主动脉瓣置换术的患者,这些患者均在一家单中心三级学术医院接受了 SAPIEN 3 经导管心脏瓣膜(爱德华兹生命科学公司)治疗,并同时使用了 ClearSight 装置和有创桡动脉压力线进行监测。患者于 2019 年 10 月至 12 月接受了经导管主动脉瓣置换术。

地点

单中心三级学术医学中心。

参与者

该研究共纳入 20 例患者,共 2243 个独特的血压数据点。

干预措施

无。

测量和主要结果

有创桡动脉压力线与 ClearSight 装置之间观察到收缩压(相关系数 0.86;p < 0.001)、舒张压(相关系数 0.56;p < 0.001)和平均动脉压(相关系数 0.78;p < 0.001)之间存在统计学显著相关性。 Bland-Altman 分析用于评估两种方法之间收缩压、舒张压和平均动脉压的一致性。动脉线与 ClearSight 装置之间收缩压的结果如下:偏差=9.8 ± 10.1,百分比偏差=7.6%,平均误差=15.8%。动脉线与 ClearSight 装置之间舒张压的结果如下:偏差=-5.9 ± 7.8,百分比偏差=10.7%,平均误差=28.4%。动脉线与 ClearSight 装置之间平均动脉压的结果如下:偏差=0.3 ± 7.4,百分比偏差=0.4%,平均误差=18.3%。收缩压、舒张压和平均动脉压的一致性率分别为 100%、95.1%和 98.8%。

结论

对于接受择期经导管主动脉瓣置换术的严重主动脉瓣狭窄患者,ClearSight 装置在测量平均动脉压、收缩压和舒张压方面的准确性、一致性和精密度令人信服。