Finite element analysis of the influence of the posterior tibial slope on mobile-bearing unicompartmental knee arthroplasty.

机构信息

Joint Reconstruction Center, Department of Orthopaedic Surgery, Yonsei Sarang Hospital, 10 Hyoryeong-ro, Seocho-gu, Seoul 06698, Republic of Korea.

Department of Mechanical Engineering, Yonsei University, 50 Yonsei-ro, Seodaemun-gu, Seoul 03722, Republic of Korea.

出版信息

Knee. 2021 Mar;29:116-125. doi: 10.1016/j.knee.2021.01.004. Epub 2021 Feb 18.

Abstract

BACKGROUND

The most common modes of failure reported in unicompartmental knee arthroplasty (UKA) in its first two decades were wear on the polyethylene (PE) insert, component loosening, and progressive osteoarthritis in the other compartment. The rates of implant failure due to poor component positioning in patients who have undergone UKA have been reported. However, the effect of the posterior tibial slope on the biomechanical behavior of mobile-bearing Oxford medial UKA remains unknown.

METHODS

We applied finite element (FE) analysis to evaluate the effects of the posterior tibial slope in mobile-bearing UKA on the contact stresses in the superior and inferior surfaces of PE inserts and articular cartilage as well as the forces exerted on the anterior cruciate ligament (ACL). Seven FE models for posterior tibial slopes of -1°, 1°, 3°, 5°, 7°, 9°, and 11° were developed and analyzed under normal-level walking conditions based on this approach.

RESULTS

The maximum contact stresses on both the superior and inferior surfaces of the PE insert decreased as the posterior tibial slope increased. However, the maximum contact stress on the lateral articular cartilage and the force exerted on the ACL increased as the posterior tibial slope increased.

CONCLUSIONS

Increasing the tibial slope led to a reduction in the contact stress on the PE insert. However, a high contact stress on the other compartment and increased ACL force can cause progressive osteoarthritis in the other compartment and failure of the ACL.

摘要

背景

在单髁膝关节置换术(UKA)的头二十年中,最常见的失效模式是聚乙烯(PE)插入物的磨损、部件松动和对侧间室的进行性骨关节炎。已经报道了由于 UKA 患者植入物位置不佳导致植入物失效的发生率。然而,胫骨后倾角对活动衬垫牛津内侧 UKA 的生物力学行为的影响尚不清楚。

方法

我们应用有限元(FE)分析来评估活动衬垫 UKA 中胫骨后倾角对 PE 插入物和关节软骨上下表面接触应力以及前交叉韧带(ACL)上施加的力的影响。根据该方法,基于正常步行条件,为-1°、1°、3°、5°、7°、9°和 11°的胫骨后倾角开发并分析了七个 FE 模型。

结果

随着胫骨后倾角的增加,PE 插入物上下表面的最大接触应力均减小。然而,随着胫骨后倾角的增加,外侧关节软骨的最大接触应力和 ACL 上的力增加。

结论

增加胫骨斜率会降低 PE 插入物的接触应力。然而,其他部位的高接触应力和 ACL 力的增加会导致其他部位的进行性骨关节炎和 ACL 失效。