Department of Chemistry and the MOE Key Laboratory of Spectrochemical Analysis & Instrumentation, College of Chemistry and Chemical Engineering, Xiamen University, Xiamen 361005, China.
Department of Chemistry and Applied Biosciences, ETH Zurich, CH-8093 Zurich , Switzerland.
Anal Chem. 2021 Apr 20;93(15):6232-6238. doi: 10.1021/acs.analchem.1c00549. Epub 2021 Apr 7.
Mass spectrometry imaging (MSI) has become a powerful tool in diverse fields, for example, life science, biomaterials, and catalysis, for its ability of and real-time visualization of the location of chemical compounds in samples. Although laser ablation (LA) achieves high spatial resolution in MSI, the ion yield can be very low. We therefore combined an LA system with an ambient ion source for post-ionization and an atmospheric pressure (AP) inlet mass spectrometer to construct a novel AP-MSI platform. A dielectric barrier discharge ionization (DBDI) source is operated in the "active sampling capillary" configuration, can be coupled to any mass spectrometer with an AP interface, and possesses high ion transmission efficiency. This study presents some application examples based on LA-DBDI, a low-cost and flexible strategy for AP-MSI, which does not require any sample pretreatment, and we show MS imaging of endogenous species in a traditional Chinese herbal medicine and of a drug molecule in zebra fish tissue, with a lateral resolution of ≈20 μm.
质谱成像(MSI)已成为生命科学、生物材料和催化等多个领域的有力工具,因为它能够实时可视化样品中化合物的位置。尽管激光烧蚀(LA)在 MSI 中实现了高空间分辨率,但离子产率可能非常低。因此,我们将 LA 系统与环境离子源进行后电离以及大气压(AP)进样质谱仪相结合,构建了一种新型的 AP-MSI 平台。介电阻挡放电电离(DBDI)源以“主动采样毛细管”的配置运行,可与任何具有 AP 接口的质谱仪耦合,并且具有高的离子传输效率。本研究基于 LA-DBDI 提供了一些应用实例,LA-DBDI 是一种低成本且灵活的 AP-MSI 策略,不需要任何样品预处理,我们展示了传统中药中内源性物质和斑马鱼组织中药物分子的 MS 成像,横向分辨率约为 20 μm。