一款用于植入式系统的、具备同时近场功率传输和数据遥测功能的27Mbps、0.08毫米CMOS收发器。

A 27-Mbps, 0.08-mm CMOS Transceiver with Simultaneous Near-field Power Transmission and Data Telemetry for Implantable Systems.

作者信息

Thimot Jordan, Kim Kukjoo, Shi Chen, Shepard Kenneth L

机构信息

Bioelectronic Systems Lab, Department of Electrical Engineering, Columbia University, New York NY.

出版信息

Proc Cust Integr Circuits Conf. 2020 Mar;2020. doi: 10.1109/CICC48029.2020.9075888. Epub 2020 Apr 23.

Abstract

This paper describes an inductively powered 27-Mbps, 0.08-mm CMOS transceiver with integrated RF receiver coils for simultaneous two-way, near-field data telemetry and power transmission for implantable systems. A four-coil inductive link operates at a 27-MHz carrier for power and a 700-MHz carrier for data telemetry with the antennae taking an area of only 2 mm by 2 mm. Amplitude-shift-keying (ASK) modulation is used for data downlink at 6.6 kbps and load-shift keying (LSK) backscattering is used for data uplink at 27 Mbps. The transceiver consumes 2.7 mW and can power a load consuming up to an additional 1.5 mW. Implemented in a 0.18-um silicon-on-insulator (SOI) technology, post-processing steps are used to decrease chip thickness to approximately 15um, making the chip flexible with a tissue-like form factor and removing the effects of the substrate on coil performance. Power harvesting circuitry, including passive rectifier, voltage regulator, RF limiter, ASK and LSK modulator, clock generator, and digital controller are positioned adjacent to the coils and limited to an area of 0.5 mm by 2mm. Complete transceiver functionality of the system has been achieved with overall power transfer efficiency (PTE) of 1.04% through 1 mm of tissue phantom between reader and implant.

摘要

本文介绍了一种采用电感供电的27Mbps、0.08mm CMOS收发器,其集成了射频接收线圈,用于植入式系统的同时双向近场数据遥测和电力传输。一种四线圈电感链路在27MHz载波上用于供电,在700MHz载波上用于数据遥测,天线占用面积仅为2mm×2mm。幅度键控(ASK)调制用于6.6kbps的数据下行链路,负载移相键控(LSK)反向散射用于27Mbps的数据上行链路。该收发器功耗为2.7mW,可为额外功耗高达1.5mW的负载供电。采用0.18μm绝缘体上硅(SOI)技术实现,后处理步骤用于将芯片厚度减小到约15μm,使芯片具有类似组织的外形且具有灵活性,并消除衬底对线圈性能的影响。功率收集电路,包括无源整流器、稳压器、射频限制器、ASK和LSK调制器、时钟发生器以及数字控制器,位于线圈附近,面积限制在0.5mm×2mm。通过读取器和植入物之间1mm厚的组织模型,系统实现了完整的收发器功能,总功率传输效率(PTE)为1.04%。

相似文献

1
A 27-Mbps, 0.08-mm CMOS Transceiver with Simultaneous Near-field Power Transmission and Data Telemetry for Implantable Systems.
Proc Cust Integr Circuits Conf. 2020 Mar;2020. doi: 10.1109/CICC48029.2020.9075888. Epub 2020 Apr 23.
2
An Integrated Passive Phase-Shift Keying Modulator for Biomedical Implants With Power Telemetry Over a Single Inductive Link.
IEEE Trans Biomed Circuits Syst. 2017 Feb;11(1):64-77. doi: 10.1109/TBCAS.2016.2580513. Epub 2016 Sep 15.
3
A Magnetic-Balanced Inductive Link for the Simultaneous Uplink Data and Power Telemetry.
Sensors (Basel). 2017 Aug 2;17(8):1768. doi: 10.3390/s17081768.
4
An NFC on Two-Coil WPT Link for Implantable Biomedical Sensors under Ultra-Weak Coupling.
Sensors (Basel). 2017 Jun 11;17(6):1358. doi: 10.3390/s17061358.
5
A Single-Chip Full-Duplex High Speed Transceiver for Multi-Site Stimulating and Recording Neural Implants.
IEEE Trans Biomed Circuits Syst. 2016 Jun;10(3):643-53. doi: 10.1109/TBCAS.2015.2466592. Epub 2015 Oct 12.
6
An 11 μW Sub-pJ/bit Reconfigurable Transceiver for mm-Sized Wireless Implants.
IEEE Trans Biomed Circuits Syst. 2016 Feb;10(1):175-85. doi: 10.1109/TBCAS.2014.2371031. Epub 2015 Jan 20.
7
A 13.56-mbps pulse delay modulation based transceiver for simultaneous near-field data and power transmission.
IEEE Trans Biomed Circuits Syst. 2015 Feb;9(1):1-11. doi: 10.1109/TBCAS.2014.2304956. Epub 2014 Apr 17.
8
A 2.4 GHz ULP OOK Single-Chip Transceiver for Healthcare Applications.
IEEE Trans Biomed Circuits Syst. 2011 Dec;5(6):523-34. doi: 10.1109/TBCAS.2011.2173340.
10
A 11 mW 2.4 GHz 0.18 µm CMOS Transceivers for Wireless Sensor Networks.
Sensors (Basel). 2017 Jan 24;17(2):223. doi: 10.3390/s17020223.

引用本文的文献

1
Millimeter-scale radioluminescent power for electronic sensors.
iScience. 2024 Dec 25;28(1):111686. doi: 10.1016/j.isci.2024.111686. eCollection 2025 Jan 17.
2
An Energy-Efficient and High-Data-Rate IR-UWB Transmitter for Intracortical Neural Sensing Interfaces.
IEEE J Solid-State Circuits. 2022 Dec;57(12):3656-3668. doi: 10.1109/JSSC.2022.3212672. Epub 2022 Oct 18.

本文引用的文献

1
An Integrated Passive Phase-Shift Keying Modulator for Biomedical Implants With Power Telemetry Over a Single Inductive Link.
IEEE Trans Biomed Circuits Syst. 2017 Feb;11(1):64-77. doi: 10.1109/TBCAS.2016.2580513. Epub 2016 Sep 15.
2
30 pJ/b, 67 Mbps, Centimeter-to-Meter Range Data Telemetry With an IR-UWB Wireless Link.
IEEE Trans Biomed Circuits Syst. 2015 Jun;9(3):362-9. doi: 10.1109/TBCAS.2014.2328492. Epub 2014 Aug 12.
3
A 13.56-mbps pulse delay modulation based transceiver for simultaneous near-field data and power transmission.
IEEE Trans Biomed Circuits Syst. 2015 Feb;9(1):1-11. doi: 10.1109/TBCAS.2014.2304956. Epub 2014 Apr 17.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验