基于当前临床实践的125I与103Pd协作性眼黑色素瘤研究(COMS)斑块的视盘剂量比较

Optic Disc Dose Comparison Between 125I and 103Pd Collaborative Ocular Melanoma Study (COMS) Plaques Based on Current Clinical Practice.

作者信息

Lee Yongsook C, Lin Shih-Chi, Kim Yongbok

机构信息

Radiation Oncology, Miami Cancer Institute, Miami, USA.

Radiation Oncology, Norwalk Hospital, Norwalk, USA.

出版信息

Cureus. 2021 Jun 28;13(6):e15980. doi: 10.7759/cureus.15980. eCollection 2021 Jun.

Abstract

Purpose The purpose of this study is to compare optic disc dose (ODD) between I and Pd Collaborative Ocular Melanoma Study (COMS) plaques in ocular brachytherapy. Methods A previously validated in-house brachytherapy dose calculation program was used for ODD calculations. ODD was calculated as a function of tumor margin-to-optic disc distance (DT) up to 5 mm for various tumor basal dimensions (BDs), for a prescription depth of 5 mm, and for standard and notched COMS plaques loaded with I (model: IAI-125A) and Pd (model: IAPd-103A) seeds. ODD calculations were repeated for prescription depths from 2 mm to 10 mm in 1 mm intervals. A prescribed dose of 85 Gy (irradiation time: 120 hours) was normalized to each prescription depth. Dose conversion factors (DCFs) for each prescription depth were calculated by taking a ratio of [total reference air kerma (TRAK) per seed]  to [TRAK per seed]. ODD reduction by notched COMS plaques was calculated for each prescription depth by subtracting ODD for notched COMS plaques from ODD for standard COMS plaques. Results Trends of ODD as a function of DT for various BDs are similar between the two seed types in both standard and notched COMS plaques. However, due to the energy difference, there exists a transition distance (d) for each BD in each plaque at which ODD for I COMS plaques equals that for Pd COMS plaques. For small BDs, at DT <d, ODD for Pd COMS plaques is higher than that for I COMS plaques while at DT >d, the opposite is observed. For the largest 1-3 BD(s), contrarily, d occurs within the tumor, and thus, ODD for I COMS plaquesis always higher. Trends of ODD reduction by notched COMS plaques as a function of DT for various BDs are the same for the two seed types except that maximum ODD reduction by Pd COMS notched plaques is larger. DCF increases with increasing prescription depth for both seed types. Conclusions There exist ODD differences between I and Pd COMS plaques and the differences depend on DT, BD, plaque size, and prescription depth.

摘要

目的 本研究旨在比较眼近距离放射治疗中碘-125(I)和钯-103(Pd)协同眼黑色素瘤研究(COMS)敷贴器的视盘剂量(ODD)。方法 使用先前经验证的内部近距离放射治疗剂量计算程序进行ODD计算。对于各种肿瘤基底尺寸(BD)、处方深度为5 mm以及装载I(型号:IAI-125A)和Pd(型号:IAPd-103A)籽源的标准和带缺口COMS敷贴器,计算肿瘤边缘至视盘距离(DT)达5 mm时的ODD作为其函数。以1 mm间隔对2 mm至10 mm的处方深度重复进行ODD计算。将规定剂量85 Gy(照射时间:120小时)归一化至每个处方深度。通过取[每个籽源的总参考空气比释动能(TRAK)]与[每个籽源的TRAK]之比,计算每个处方深度的剂量转换因子(DCF)。通过从标准COMS敷贴器的ODD中减去带缺口COMS敷贴器的ODD,计算每个处方深度带缺口COMS敷贴器导致的ODD降低。结果 在标准和带缺口COMS敷贴器中,两种籽源类型中ODD随DT变化的趋势对于各种BD而言相似。然而,由于能量差异,每个敷贴器中每种BD都存在一个转变距离(d),在该距离处I COMS敷贴器的ODD等于Pd COMS敷贴器的ODD。对于小BD,在DT <d时,Pd COMS敷贴器的ODD高于I COMS敷贴器,而在DT >d时,情况相反。相反,对于最大的1 - 3种BD,d出现在肿瘤内部,因此,I COMS敷贴器的ODD始终更高。两种籽源类型中,带缺口COMS敷贴器导致的ODD降低随DT变化的趋势对于各种BD而言相同,只是Pd带缺口敷贴器的最大ODD降低更大。两种籽源类型的DCF均随处方深度增加而增加。结论 I和Pd COMS敷贴器之间存在ODD差异,且这些差异取决于DT、BD、敷贴器尺寸和处方深度。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/d95e/8317005/d895fdfe63ac/cureus-0013-00000015980-i01.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验