Liverpool School of Tropical Medicine, Centre for Drugs and Diagnostics, Liverpool, L3 5QA, UK.
Departments of Vector Biology and Tropical Disease Biology, Liverpool School of Tropical Medicine, Centre for Neglected Tropical Diseases, Liverpool, L3 5QA, UK.
Sci Rep. 2021 Sep 15;11(1):18313. doi: 10.1038/s41598-021-97489-9.
In the context of the coronavirus disease 2019 (COVID-19) pandemic there has been an increase of the use of antigen-detection rapid diagnostic tests (Ag-RDT). The performance of Ag-RDT vary greatly between manufacturers and evaluating their analytical limit of detection (LOD) has become high priority. Here we describe a manufacturer-independent evaluation of the LOD of 19 marketed Ag-RDT using live SARS-CoV-2 spiked in different matrices: direct culture supernatant, a dry swab, and a swab in Amies. Additionally, the LOD using dry swab was investigated after 7 days' storage at - 80 °C of the SARS-CoV-2 serial dilutions. An LOD of ≈ 5.0 × 10 pfu/ml (1.0 × 10 genome copies/ml) in culture media is defined as acceptable by the World Health Organization. Fourteen of 19 Ag-RDTs (ActiveXpress, Espline, Excalibur, Innova, Joysbio, Mologic, NowCheck, Orient, PanBio, RespiStrip, Roche, Standard-F, Standard-Q and Sure-Status) exceeded this performance criteria using direct culture supernatant applied to the Ag-RDT. Six Ag-RDT were not compatible with Amies media and a decreased sensitivity of 2 to 20-fold was observed for eleven tests on the stored dilutions at - 80 °C for 7 days. Here, we provide analytical sensitivity data to guide appropriate test and sample type selection for use and for future Ag-RDT evaluations.
在 2019 年冠状病毒病(COVID-19)大流行背景下,抗原检测快速诊断检测(Ag-RDT)的使用有所增加。Ag-RDT 的性能在制造商之间差异很大,评估其分析检测限(LOD)已成为当务之急。在这里,我们使用活 SARS-CoV-2 对 19 种市售 Ag-RDT 的 LOD 进行了制造商独立评估,该病毒分别掺入不同基质:直接培养上清液、干燥拭子和 Amies 拭子。此外,还研究了在 -80°C 下储存 7 天后 SARS-CoV-2 系列稀释液干燥拭子的 LOD。世界卫生组织(WHO)定义在培养基中约为 5.0×10 pfu/ml(1.0×10 基因组拷贝/ml)的 LOD 为可接受水平。在直接培养上清液应用于 Ag-RDT 时,19 种 Ag-RDT 中的 14 种(ActiveXpress、Espline、Excalibur、Innova、Joysbio、Mologic、NowCheck、Orient、PanBio、RespiStrip、罗氏、Standard-F、Standard-Q 和 Sure-Status)超过了此性能标准。六种 Ag-RDT 与 Amies 介质不兼容,并且在 -80°C 下储存 7 天后,对 11 种测试的敏感性降低了 2 至 20 倍。在这里,我们提供了分析灵敏度数据,以指导适当的测试和样品类型选择,以用于检测和未来的 Ag-RDT 评估。