Department of Laboratory Medicine, Zhongshan Hospital, Fudan University, Shanghai, China.
Department of Laboratory Medicine, Xiamen Branch, Zhongshan Hospital, Fudan University, Xiamen, China.
J Clin Lab Anal. 2021 Nov;35(11):e23834. doi: 10.1002/jcla.23834. Epub 2021 Oct 12.
Flow cytometry is a powerful technique that provides information regarding cell properties. In this study, we evaluated the analytical performance of a new flow cytometer, the 10-color BD FACSLyric , which could help doctors obtain reliable test results prior to clinical research.
We used Sphero Rainbow Calibration Particles and the Sphero Nano Fluorescent Particle Size Standard Kit to validate the fluorescence sensitivity and linearity. The Beckman Coulter IMMUNO-TROL Cell was used as the quality control to evaluate the accuracy and reproducibility of surface markers detected by the flow cytometer. Furthermore, BD Calibrate APC Beads and CS&T Research Beads were applied to calculate the carry-over contamination rate and assess the instrument stability.
A linear regression equation between the molecules of equivalent soluble fluorochrome and fluorescence detection limit showed a good linear fit (R > 0.99). The minimum bead size detected by side scatter was 0.22 μm. The coefficient of variation percentage of each fluorescence channel was below 2%, and the carry-over contamination rate of the cytometer was under 0.2%. After running the BD FACSLyric cytometer continuously for 8 h, the median fluorescence index of particles remained close to that at the time of cytometer startup.
The 10-color BD FACSLyric cytometer showed good performance in the evaluation performed in this study and may be trusted to provide accurate results for clinical research.
流式细胞术是一种强大的技术,可提供有关细胞特性的信息。在这项研究中,我们评估了一种新型流式细胞仪 BD FACSLyric 的分析性能,这有助于医生在临床研究之前获得可靠的测试结果。
我们使用 Sphero Rainbow 校准粒子和 Sphero 纳米荧光粒度标准试剂盒来验证荧光灵敏度和线性度。使用贝克曼库尔特 IMMUNO-TROL 细胞作为质量控制,以评估流式细胞仪检测到的表面标志物的准确性和重现性。此外,使用 BD Calibrate APC Beads 和 CS&T Research Beads 计算携带污染率并评估仪器稳定性。
等效可溶性荧光染料分子与荧光检测限之间的线性回归方程显示出良好的线性拟合(R > 0.99)。通过侧散射检测到的最小珠粒尺寸为 0.22 μm。每个荧光通道的变异系数百分比均低于 2%,并且仪器的携带污染率低于 0.2%。在连续运行 BD FACSLyric 流式细胞仪 8 小时后,颗粒的中位荧光指数仍接近流式细胞仪启动时的荧光指数。
在本研究中进行的评估中,10 色 BD FACSLyric 流式细胞仪表现出良好的性能,可信赖地为临床研究提供准确的结果。