使用三维运动分析仪对一种用于腕关节疾病的新型人工肌肉型动态矫形器的分析。

Analysis of a New Artificial Muscle Type Dynamic Orthosis for Wrist Joint Disease Using a Three-dimensional Motion Analyzer.

作者信息

Nakayama Jun, Sunagawa Kosaku, Ogawa Kazunori, Oka Hisao

机构信息

Department of Allied Health Sciences, Kansai University of Welfare Sciences, Osaka, Japan.

Daiya Industrial Inc, Okayama City, Japan.

出版信息

Prog Rehabil Med. 2021 Nov 5;6:20210043. doi: 10.2490/prm.20210043. eCollection 2021.

Abstract

OBJECTIVES

Recently, the dart-throwing motion (DTM) has attracted attention as a functional direction of wrist joint motion. Consequently, we devised a new artificial muscle-type dynamic orthosis (Dynamic Traction Splint by Artificial Muscle, DTSaM) to reproduce DTM. This study analyzed the automatic motion of the wrist joint using a three-dimensional motion analysis system to assess how closely the DTSaM replicates DTM.

METHODS

The DTSaM orthosis incorporates two McKibben-type rubber artificial muscles, and measurements were performed using image analysis software and a three-dimensional motion analysis system. The wrist radial angle (WRA) was defined as the angle between the line connecting the head of the index finger metacarpal to the radial styloid process and the line connecting the radial styloid process to the lateral epicondyle of the humerus. WRAs were investigated from 60° of palmar flexion to 60° of dorsiflexion.

RESULTS

For dorsiflexion, comparisons of radial deviation and wrist ulnar angle (WUA) between the motion obtained using the DTSaM orthosis and active DTM showed a significant difference (P <0.05) at 30° and 60°, respectively. For palmar flexion, the same comparison showed a significant difference for both the ulnar and radial deviations (30°, P <0.05; and 60°, P <0.01, respectively). Furthermore, WUA showed a significant difference at 50° (P <0.05). Intraclass correlation coefficient analyses yielded good reliability with an average value of ≥0.8.

CONCLUSION

The DTSaM orthosis produces a motion similar to the DTM. It is hoped that the use of the DTSaM orthosis will help to shorten the treatment period for patients with wrist disease.

摘要

目的

近来,投掷动作(DTM)作为腕关节运动的一种功能方向受到关注。因此,我们设计了一种新型人工肌肉型动态矫形器(人工肌肉动态牵引夹板,DTSaM)来重现DTM。本研究使用三维运动分析系统分析腕关节的自动运动,以评估DTSaM对DTM的复制程度。

方法

DTSaM矫形器包含两个麦基本型橡胶人工肌肉,并使用图像分析软件和三维运动分析系统进行测量。腕桡角(WRA)定义为连接食指掌骨头与桡骨茎突的线和连接桡骨茎突与肱骨外侧髁的线之间的夹角。研究了从掌屈60°到背屈60°的WRA。

结果

对于背屈,使用DTSaM矫形器获得的运动与主动DTM之间,在30°和60°时,桡偏和腕尺角(WUA)的比较分别显示出显著差异(P<0.05)。对于掌屈,相同比较显示尺偏和桡偏均存在显著差异(分别为30°,P<0.05;60°,P<0.01)。此外,WUA在50°时显示出显著差异(P<0.05)。组内相关系数分析得出良好的可靠性,平均值≥0.8。

结论

DTSaM矫形器产生的运动与DTM相似。希望使用DTSaM矫形器将有助于缩短腕部疾病患者的治疗周期。