Abstract

BACKGROUND

The diagnosis of hypercortisolism requires multiple biochemical investigations, due to variations in cortisol production during the 24-hour circadian cycle. Midnight serum cortisol is difficult to interpret since the threshold value is dependent on the analytical method used and is often not provided by the manufacturer. Second-generation assays are more specific than first-generation assays and may have lower threshold values.

OBJECTIVES

The aim of this study was to determine a novel threshold value of midnight serum cortisol for the biochemical diagnosis of hypercortisolism, using the Roche Cobas Cortisol® second-generation assay.

METHODS

This study was performed in adult patients hospitalized in the endocrinology unit of a university hospital. Patients had a complete assessment of their 24-hour cortisol cycle, i.e., a serum cortisol test every four hours and at least two first-line tests: late night salivary cortisol, dexamethasone suppression test and/or 24-hour urinary free cortisol. First-line tests were used to identify patients with hypercortisolism. Serum samples were analyzed by second-generation electrochemiluminescence immunoassays (ECLIA) from Roche Cobas Cortisol®.

RESULTS

Midnight serum cortisol samples were obtained from 175 hospitalized patients. The novel threshold value obtained was 157 nmol/L with a sensitivity of 82.9% (68.6 to 94.3%) and a specificity of 90.0% (85.0 to 95.0%).

CONCLUSION

Our study confirms that the threshold value of midnight serum cortisol is not comparable between first- and second-generation Roche Cobas Cortisol® assays and that the threshold value is lower with the second-generation assay.

摘要

背景

由于皮质醇在24小时昼夜周期中的分泌存在变化,高皮质醇血症的诊断需要多项生化检查。午夜血清皮质醇难以解读,因为其阈值取决于所使用的分析方法,且制造商通常未提供该阈值。第二代检测方法比第一代检测方法更具特异性,且可能具有更低的阈值。

目的

本研究旨在使用罗氏Cobas Cortisol®第二代检测方法确定用于高皮质醇血症生化诊断的午夜血清皮质醇新阈值。

方法

本研究在一家大学医院内分泌科住院的成年患者中进行。患者对其24小时皮质醇周期进行了全面评估,即每四小时进行一次血清皮质醇检测,并至少进行两项一线检测:深夜唾液皮质醇、地塞米松抑制试验和/或24小时尿游离皮质醇。一线检测用于识别高皮质醇血症患者。血清样本采用罗氏Cobas Cortisol®第二代电化学发光免疫分析(ECLIA)进行分析。

结果

从175名住院患者中获取了午夜血清皮质醇样本。获得的新阈值为157 nmol/L,灵敏度为82.9%(68.6%至94.3%),特异性为90.0%(85.0%至95.0%)。

结论

我们的研究证实,第一代和第二代罗氏Cobas Cortisol®检测方法的午夜血清皮质醇阈值不可比,且第二代检测方法的阈值更低。