Narahara K A, Thompson C J, Maublant J C, Criley J M, Mena I
J Nucl Med. 1987 Aug;28(8):1315-21.
A new automated edge detection program has been developed to estimate left ventricular mass from single photon emission computed tomographic (SPECT) 201Tl images and 14 dogs were studied. Six of the 14 dogs underwent imaging before and 5 hr after coronary artery occlusion with a closed-chest technique. True left ventricular mass was determined at time of killing within 1 hr of the last 201Tl study. Left ventricular mass determined by tomography correlated well with autopsy left ventricular mass (r = 0.94; p less than 0.001, s.e.e. = 5.9 g) over a range of 62-156 g. The intraobserver variation between repeated measurements of the same SPECT study yielded an r = 0.99; p less than 0.0001; s.e.e. = 2.3 g. The reproducibility of the mass determination was assessed in four animals with two studies performed 10-14 days apart. The estimate of left ventricular mass from the two studies were highly correlated (r = 0.98; p less than 0.001) with a mean absolute difference of 4 g (3.3%). In the six dogs with a control and postinfarct study the mean total left ventricular mass by 201Tl tomography varied by less than 3.8% (r = 0.89; p less than 0.001). In conclusion, tomographic imaging with 201Tl can define left ventricular mass accurately and reproducibly in the dog model. The ability of this 201Tl tomographic left ventricular mass program to measure both normal and infarcted tissue accurately suggests the possibility of documenting interventions designed to alter left ventricular mass and of sizing acute infarcts and assessing interventions that may alter acute infarct size.
已开发出一种新的自动边缘检测程序,用于从单光子发射计算机断层扫描(SPECT)铊-201图像估计左心室质量,并对14只狗进行了研究。14只狗中的6只采用闭胸技术在冠状动脉闭塞前和闭塞后5小时进行成像。在最后一次铊-201研究后1小时内处死动物时测定真实的左心室质量。在62 - 156克的范围内,通过断层扫描测定的左心室质量与尸检左心室质量相关性良好(r = 0.94;p < 0.001,标准误 = 5.9克)。同一SPECT研究重复测量之间的观察者内变异得出r = 0.99;p < 0.0001;标准误 = 2.3克。在4只动物中评估了质量测定的可重复性,两项研究间隔10 - 14天进行。两项研究中左心室质量的估计高度相关(r = 0.98;p < 0.001),平均绝对差异为4克(3.3%)。在6只进行了对照和梗死后研究的狗中,通过铊-201断层扫描测定的左心室总质量平均变化小于3.8%(r = 0.89;p < 0.001)。总之,铊-201断层成像能够在狗模型中准确且可重复地定义左心室质量。这种铊-201断层左心室质量程序准确测量正常和梗死组织的能力表明,有可能记录旨在改变左心室质量的干预措施,以及对急性梗死灶进行大小测量并评估可能改变急性梗死灶大小的干预措施。