Suppr超能文献

3T临床MRI系统中电悬浮和接地PET插入物的射频透明度研究。

Study on the radiofrequency transparency of electrically floating and ground PET inserts in a 3 T clinical MRI system.

作者信息

Akram Md Shahadat Hossain, Obata Takayuki, Nishikido Fumihiko, Yamaya Taiga

机构信息

Department of Advanced Nuclear Medicine Sciences, Institute for Quantum Medical Sciences, National Institutes for Quantum Science and Technology (QST), Anagawa, Inage, Chiba, Japan.

Department of Applied MRI Research, National Institute of Radiological Sciences in the National Institutes for Quantum and Radiological Science and Technology (NIRS-QST), Anagawa, Inage, Chiba, Japan.

出版信息

Med Phys. 2022 May;49(5):2965-2978. doi: 10.1002/mp.15588. Epub 2022 Mar 15.

Abstract

PURPOSE

The positron emission tomography (PET) insert for a magnetic resonance imaging (MRI) system that implements the radiofrequency (RF) built-in body coil of the MRI system as a transmitter is designed to be RF-transparent, as the coil resides outside the RF-shielded PET ring. This approach reduces the design complexities (e.g., large PET ring diameter) related to implementing a transmit coil inside the PET ring. However, achieving the required field transmission into the imaging region of interest (ROI) becomes challenging because of the RF shield of the PET insert. In this study, a modularly RF-shielded PET insert is used to investigate the RF transparency considering two electrical configurations of the RF shield, namely the electrical floating and ground configurations. The purpose is to find the differences, advantages and disadvantages of these two configurations.

METHODS

Eight copper-shielded PET detector modules (intermodular gap: 3 mm) were oriented cylindrically with an inner diameter of 234 mm. Each PET module included four-layer Lutetium-yttrium oxyorthosilicate scintillation crystal blocks and front-end readout electronics. RF-shielded twisted-pair cables were used to connect the front-end electronics with the power sources and PET data acquisition systems located outside the MRI room. In the ground configuration, both the detector and cable shields were connected to the RF ground of the MRI system. In the floating configuration, only the RF shields of the PET modules were isolated from the RF ground. Experiments were conducted using two cylindrical homogeneous phantoms in a 3 T clinical MRI system, in which the built-in body RF coil (a cylindrical volume coil of diameter 700 mm and length 540 mm) was implemented as a transceiver.

RESULTS

For both PET configurations, the RF and MR imaging performances were lower than those for the MRI-only case, and the MRI system provided specific absorption ratio (SAR) values that were almost double. The RF homogeneity and field strength, and the signal-to-noise ratio (SNR) of the MR images were mostly higher for the floating PET configuration than they were for the ground PET configuration. However, for a shorter axial field-of-view (FOV) of 125 mm, both configurations offered almost the same performance with high RF homogeneities (e.g., 76 ± 10%). Moreover, for both PET configurations, 56 ± 6% larger RF pulse amplitudes were required for MR imaging purposes. The increased power is mostly absorbed in the conductive shields in the form of shielding RF eddy currents; as a result, the SAR values only in the phantoms were estimated to be close to the MRI-only values.

CONCLUSIONS

The floating PET configuration showed higher RF transparency under all experimental setups. For a relatively short axial FOV of 125 mm, the ground configuration also performed well which indicated that an RF-penetrable PET insert with the conventional design (e.g., the ground configuration) might also become possible. However, some design modifications (e.g., a wider intermodular gap and using the RF receiver coil inside the PET insert) should improve the RF performance to the level of the MRI-only case.

摘要

目的

用于磁共振成像(MRI)系统的正电子发射断层扫描(PET)插入件将MRI系统的射频(RF)内置体线圈用作发射器,由于该线圈位于RF屏蔽的PET环外部,因此设计为RF透明。这种方法减少了与在PET环内实施发射线圈相关的设计复杂性(例如,大的PET环直径)。然而,由于PET插入件的RF屏蔽,实现所需的场传输到感兴趣的成像区域(ROI)变得具有挑战性。在本研究中,使用模块化RF屏蔽的PET插入件来研究考虑RF屏蔽的两种电气配置(即电浮动和接地配置)的RF透明度。目的是找出这两种配置的差异、优点和缺点。

方法

八个铜屏蔽的PET探测器模块(模块间间隙:3毫米)以圆柱形排列,内径为234毫米。每个PET模块包括四层正硅酸镥钇闪烁晶体块和前端读出电子设备。使用RF屏蔽双绞线电缆将前端电子设备与位于MRI室外部的电源和PET数据采集系统连接。在接地配置中,探测器和电缆屏蔽都连接到MRI系统的RF地。在浮动配置中,仅PET模块的RF屏蔽与RF地隔离。在3T临床MRI系统中使用两个圆柱形均匀体模进行实验,其中内置体RF线圈(直径700毫米、长度540毫米的圆柱形容积线圈)用作收发器。

结果

对于两种PET配置,RF和MR成像性能均低于仅进行MRI的情况,并且MRI系统提供的比吸收率(SAR)值几乎翻倍。对于浮动PET配置,MR图像的RF均匀性和场强以及信噪比(SNR)大多高于接地PET配置。然而,对于125毫米的较短轴向视野(FOV),两种配置具有几乎相同的性能,具有高RF均匀性(例如,76±10%)。此外,对于两种PET配置,MR成像需要大56±6%的RF脉冲幅度。增加的功率主要以屏蔽RF涡流的形式被导电屏蔽吸收;因此,仅在体模中的SAR值估计接近仅进行MRI的值。

结论

在所有实验设置下,浮动PET配置显示出更高的RF透明度。对于125毫米的相对较短轴向FOV,接地配置也表现良好,这表明具有传统设计(例如,接地配置)的RF可穿透PET插入件也可能成为可能。然而,一些设计修改(例如,更宽的模块间间隙和在PET插入件内使用RF接收线圈)应将RF性能提高到仅进行MRI的水平。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验