Clinical experience with Mentor inflatable penile prosthesis in 206 patients.

作者信息

Merrill D C

出版信息

Urology. 1986 Sep;28(3):185-9. doi: 10.1016/0090-4295(86)90039-7.

Abstract

The Mentor inflatable penile prosthesis (IPP) has been implanted in 206 patients with organic (95.2%) or psychologic (4.8%) impotency. The patients have been followed up for an average of 18.1 (4-42) months. A satisfactory result has been achieved in 98 per cent of these patients. A life analysis of the series shows that 88 per cent of all devices implanted will survive forty-two months without mechanical failure. One hundred sixty-four of these patients had not had previous penile surgery while 42 had had one or more penile prostheses implanted previously. The projected forty-two-month survival for the Mentor prosthesis in these subgroups is 94 per cent and 74 per cent, respectively. The improved life expectancy of the Mentor IPP, compared with the Scott IPP, is due to the superior durability of Mentor's polyurethane cylinders and to the increased reliability of the Mentor snap-on connector system. In this respect, there have been only one cylinder failure and 3 connector failures in the Mentor series. The connector failures occurred early in the series before the development of connector pliers; there have been no connector failures in the last 202 consecutive implantations. Wear-induced tubing leaks, the most common cause of device failure, occurred in 2.9 per cent of all implantations. Tubing failures were more common in patients who have had previous surgery, occurring in 9 per cent of salvage procedures and 1.3 per cent of primary implantations. The reason for the difference in tubing survival in virgin and salvage procedures is unknown; possibly the subcutaneous scar tissue present in secondary surgical procedures increases the chance of tubing segments touching and abrasing each other. Laboratory studies suggest that the nylon-reinforced tubing used in the manufacture of devices implanted since 1985 is more durable than is the standard silicone tubing used in the construction of devices implanted previously.

摘要

相似文献

1
Clinical experience with Mentor inflatable penile prosthesis in 206 patients.
Urology. 1986 Sep;28(3):185-9. doi: 10.1016/0090-4295(86)90039-7.
2
Clinical experience with the Mentor inflatable penile prosthesis in 301 patients.
J Urol. 1988 Dec;140(6):1424-7. doi: 10.1016/s0022-5347(17)42063-5.
4
Mentor inflatable penile prosthesis.
Urology. 1983 Nov;22(5):504-5. doi: 10.1016/0090-4295(83)90229-7.
5
Mentor inflatable penile prosthesis. Preliminary clinical results in 30 patients.
Urology. 1984 May;23(5 Spec No):72-4. doi: 10.1016/0090-4295(84)90245-0.
6
Clinical experience with Scott inflatable penile prosthesis in 150 patients.
Urology. 1983 Oct;22(4):371-5. doi: 10.1016/0090-4295(83)90412-0.
7
Reliability of AMS M700 inflatable penile prosthesis.
Urology. 1986 Nov;28(5):385-7. doi: 10.1016/0090-4295(86)90067-1.
8
42 months of experience with the Mentor inflatable penile prosthesis.
J Urol. 1988 Jan;139(1):48-9. doi: 10.1016/s0022-5347(17)42286-5.
9
Scott's inflatable penile prosthesis: evaluation of mechanical survival in the series 700 model.
J Urol. 1987 Apr;137(4):676-7. doi: 10.1016/s0022-5347(17)44171-1.