Suppr超能文献

关于HeartMate 3左心室辅助装置泵内血栓形成率低的见解:人工脉冲是否能改善冲洗效果?

Insights Into the Low Rate of In-Pump Thrombosis With the HeartMate 3: Does the Artificial Pulse Improve Washout?

作者信息

Fang Peng, Du Jianjun, Boraschi Andrea, Bozzi Silvia, Redaelli Alberto, Schmid Daners Marianne, Kurtcuoglu Vartan, Consolo Filippo, de Zélicourt Diane

机构信息

School of Mechanical Engineering and Automation, Harbin Institute of Technology, Shenzhen, Shenzhen, China.

The Interface Group, Institute of Physiology, University of Zurich, Zurich, Switzerland.

出版信息

Front Cardiovasc Med. 2022 Mar 11;9:775780. doi: 10.3389/fcvm.2022.775780. eCollection 2022.

Abstract

While earlier studies reported no relevant effect of the HeartMate 3 (HM3) artificial pulse (AP) on bulk pump washout, its effect on regions with prolonged residence times remains unexplored. Using numerical simulations, we compared pump washout in the HM3 with and without AP with a focus on the clearance of the last 5% of the pump volume. Results were examined in terms of flush-volume ( , number of times the pump was flushed with new blood) to probe the effect of the AP independent of changing flow rate. Irrespective of the flow condition, the HM3 washout scaled linearly with flush volume up to 70% washout and slowed down for the last 30%. Flush volumes needed to washout 95% of the pump were comparable with and without the AP (1.3-1.4 ), while 99% washout required 2.1-2.2 with the AP vs. 2.5 without the AP. The AP enhanced washout of the bend relief and near-wall regions. It also transiently shifted or eliminated stagnation regions and led to rapid wall shear stress fluctuations below the rotor and in the secondary flow path. Our results suggest potential benefits of the AP for clearance of fluid regions that might elicit in-pump thrombosis and provide possible mechanistic rationale behind clinical data showing very low rate of in-pump thrombosis with the HM3. Further optimization of the AP sequence is warranted to balance washout efficacy while limiting blood damage.

摘要

虽然早期研究报告称HeartMate 3(HM3)人工脉冲(AP)对整体泵冲洗没有相关影响,但其对停留时间延长区域的影响仍未得到探索。我们使用数值模拟,比较了有和没有AP的HM3中的泵冲洗情况,重点关注泵容积最后5%的清除情况。根据冲洗量( ,用新鲜血液冲洗泵的次数)来检查结果,以探究AP独立于流量变化的影响。无论流动条件如何,HM3的冲洗在冲洗量达到70%之前与冲洗量呈线性关系,在最后30%时减缓。冲洗掉泵95%容积所需的冲洗量在有和没有AP时相当(1.3 - 1.4 ),而冲洗掉99%则在有AP时需要2.1 - 2.2 ,无AP时需要2.5 。AP增强了弯曲缓解区域和近壁区域的冲洗。它还会暂时移动或消除停滞区域,并导致转子下方和二次流动路径中的壁面剪应力快速波动。我们的结果表明,AP对于清除可能引发泵内血栓形成的流体区域具有潜在益处,并为显示HM3泵内血栓形成率极低的临床数据提供了可能的机制依据。有必要进一步优化AP序列,以在限制血液损伤的同时平衡冲洗效果。