Li Xinhua, Steigerwalt David, Rehani Madan M
Massachusetts General Hospital, 55 Fruit Str, Boston, MA 02114, USA.
Imalogix Plc, Bryn Mawr, PA 19010, USA.
Eur J Radiol. 2022 Jun;151:110289. doi: 10.1016/j.ejrad.2022.110289. Epub 2022 Apr 4.
To examine the impact of patient size on dose indices and develop size-based reference levels (50th and 75th percentiles) for 20 body CT exams for routine and organ-specific clinical indications.
Based on effective diameter estimated from adult body CT, each acquisition was classified into T-shirt size as XXS, XS, S, M, L, XL, and XXL. Radiation dose indices for each size and each exam type were correlated.
About 0.93 million CT exams from 256 CT facilities in the United States were analysed. Taking T-shirt size M as a reference, the CTDI for other sizes were: XXS (∼60%), XS (∼65%), S (∼75%), L (∼130%), XL (∼165%), XXL (∼210%), or grossly small patients received about 60% of the dose as compared to M sized patients and XXL required doubling the dose. Taking ratio of the dose indices of the largest to smallest size, it was evident that SSDE variation was much less (about 50%) than that in CTDI, but there was still nearly 40 to 220% variation in SSDE across the range of t-shirt sizes. The 50th and 75th percentile values are presented for CTDI, SSDE and DLP for each of the 20 CT exams and for each of the seven T-shirt sizes.
A novel approach expressing body habitus in terms of T-shirt size is not only simple and intuitive, but it also provides a tool to have a perception of differences in dose metrices among patients of different body build.
研究患者体型对剂量指标的影响,并为20种针对常规和器官特异性临床指征的身体CT检查制定基于体型的参考水平(第50和第75百分位数)。
根据成人身体CT估算的有效直径,将每次扫描分为XXS、XS、S、M、L、XL和XXL等T恤尺码。对每种尺码和每种检查类型的辐射剂量指标进行相关性分析。
分析了来自美国256个CT设备的约93万次CT检查。以T恤尺码M为参考,其他尺码的CTDI分别为:XXS(约60%)、XS(约65%)、S(约75%)、L(约130%)、XL(约165%)、XXL(约210%),即体型非常小的患者接受的剂量约为M尺码患者的60%,而XXL尺码患者所需剂量翻倍。从最大尺码与最小尺码的剂量指标比值来看,很明显SSDE的变化比CTDI小得多(约50%),但在整个T恤尺码范围内,SSDE仍有近40%至220%的变化。给出了20种CT检查中每种检查以及7种T恤尺码中每种尺码的CTDI、SSDE和DLP的第50和第75百分位数。
一种用T恤尺码来表示体型的新方法不仅简单直观, 还提供了一种工具,有助于了解不同体型患者之间剂量指标的差异。