J Diabetes Sci Technol. 2023 Sep;17(5):1317-1320. doi: 10.1177/19322968221095055. Epub 2022 May 19.
A data set comprising 28 533 oxygen partial pressure (pO) values gathered from nonarterialized fingertip capillary blood samples of a large diabetes population is reported. Data were gathered routinely as part of the ongoing clinical surveillance activities of a blood glucose monitoring (BGM) system manufacturer. Overall, a mean pO level of 75.3 mm Hg was recorded, with a standard deviation (SD) of 10.6 mm Hg and a range of 27.7 to 151.5 mm Hg. This data set, unique in terms of size, gathered more than 14 years at four separate clinical sites, provides a robust and comprehensive representation of an unmanipulated pO fingertip capillary sample distribution, which should be of interest for physiological reasons, but also from a medical device development perspective, in which the effect of pO on device performance may be an important design consideration.
本文报告了一个包含 28533 个氧分压(pO)值的数据集,这些值来自于大量糖尿病患者的非动脉化指尖毛细血管血样。这些数据是作为血糖监测(BGM)系统制造商持续临床监测活动的一部分常规收集的。总体而言,记录到的平均 pO 水平为 75.3mmHg,标准差(SD)为 10.6mmHg,范围为 27.7 至 151.5mmHg。该数据集在规模上是独一无二的,在四个不同的临床地点收集了超过 14 年的数据,为未经处理的指尖毛细血管 pO 样本分布提供了一个强大而全面的代表性,这无论出于生理原因还是从医疗器械开发的角度来看都应该是有趣的,因为 pO 对设备性能的影响可能是一个重要的设计考虑因素。