可穿戴光电容积脉搏波传感器在远程医疗应用中连续心率监测的准确性。

Accuracy of Wearable Photoplethysmography Sensors for Continuous Heart Rate Monitoring in Telehealth Applications.

机构信息

AIT Austrian Institute of Technology GmbH, Graz, Austria.

Graz University of Technology, Graz, Austria.

出版信息

Stud Health Technol Inform. 2022 May 16;293:205-211. doi: 10.3233/SHTI220370.

Abstract

The demand for extended care for people suffering from heart failure is omnipresent. Wearables providing continuous heart rate measurement through optical sensors are of great interest due to their ease of use without the need for medical staff and their low cost. In this study, seven wearables were tested in fifteen measurement runs, with a duration of fourteen-hour each, and compared to a reference sensor. By calculating the Pearson correlation and the root mean square error, as well as the graphical representation by a Bland Altman plot, it was found that these wearables lack sufficient accuracy and may not be suitable for medical purposes.

摘要

对于心力衰竭患者的长期护理需求是普遍存在的。由于无需医护人员且成本低廉,通过光学传感器进行连续心率测量的可穿戴设备引起了极大的关注。在这项研究中,对七种可穿戴设备进行了十五次测量,每次持续十四小时,并与参考传感器进行了比较。通过计算 Pearson 相关系数和均方根误差,以及通过 Bland-Altman 图进行图形表示,发现这些可穿戴设备的准确性不足,可能不适合用于医疗目的。