Amick Michael, Ottesen Taylor D, O'Marr Jamieson, Frenkel Mikhail Y, Callahan Brooke, Grauer Jonathan N
Department of Orthopaedics and Rehabilitation, Yale School of Medicine, 800 Howard Ave, New Haven, CT, 06510, USA.
Department of Orthopaedics and Rehabilitation, Yale School of Medicine, 800 Howard Ave, New Haven, CT, 06510, USA; Harvard Combined Orthopaedic Residency Program, 55 Fruit St, Boston, MA, 02114, USA.
Spine J. 2022 Dec;22(12):2000-2005. doi: 10.1016/j.spinee.2022.07.090. Epub 2022 Jul 16.
Pedicle screws are commonly placed with lumbar/lumbosacral fusions. Triggered electromyography (tEMG), which employs the application of electrical current between the screw and a complementary anode to determine thresholds of conduction, may be utilized to confirm the safe placement of such implants. While previous research has established clinical thresholds associated with safe screw placement, there is variability in clinical practice of anode placement which could lead to unreliable measurements.
To determine the variance in pedicle screw stimulation thresholds when using four unique anode locations (ipsilateral/contralateral and paraspinal/gluteal relative to tested pedicle screws).
Prospective cohort study. Tertiary medical center.
Twenty patients undergoing lumbar/lumbosacral fusion with pedicle screws using tEMG OUTCOME MEASURES: tEMG stimulation return values are used to assess varied anode locations and reproducibility based on anode placement.
Measurements were assessed across node placement in ipsilateral/contralateral and paraspinal/gluteal locations relative to the screw being assessed. R coefficients of correlation were determined, and variances were compared with F-tests.
A total of 94 lumbosacral pedicle screws from 20 patients were assessed. Repeatability was verified using two stimulations at each location for a subset of the screws with an R of 0.96. Comparisons between the four anode locations demonstrated R values ranging from 0.76 to 0.87. F-tests comparing thresholds between each anode site demonstrated all groups not to be statistically different.
The current study, a first-of-its-kind formal evaluation of anode location for pedicle screw tEMG testing, demonstrated very strong repeatability and strong correlation with different locations of anode placement. These results suggest that there is no need to change the side of the anode for testing of left versus right screws, further supporting that placing an anode electrode into gluteal muscle is sufficient and will avoid a sharp ground needle in the surgical field.
椎弓根螺钉常用于腰椎/腰骶部融合手术。触发式肌电图(tEMG)通过在螺钉和互补阳极之间施加电流来确定传导阈值,可用于确认此类植入物的安全放置。虽然先前的研究已经确定了与螺钉安全放置相关的临床阈值,但阳极放置的临床实践存在差异,这可能导致测量结果不可靠。
确定使用四个独特的阳极位置(相对于测试椎弓根螺钉的同侧/对侧以及椎旁/臀肌)时椎弓根螺钉刺激阈值的差异。
前瞻性队列研究。三级医疗中心。
20例接受使用tEMG的椎弓根螺钉腰椎/腰骶部融合手术的患者
tEMG刺激返回值用于评估不同的阳极位置以及基于阳极放置的可重复性。
在相对于被评估螺钉的同侧/对侧以及椎旁/臀肌位置进行节点放置的测量评估。确定相关系数R,并通过F检验比较方差。
对20例患者的94枚腰骶椎弓根螺钉进行了评估。对一部分螺钉在每个位置进行两次刺激以验证重复性,R值为0.96。四个阳极位置之间的比较显示R值范围为0.76至0.87。比较每个阳极部位之间阈值的F检验表明所有组在统计学上无差异。
本研究是对椎弓根螺钉tEMG测试阳极位置的首次正式评估,显示出非常强的可重复性以及与不同阳极放置位置的强相关性。这些结果表明,在测试左右螺钉时无需改变阳极的一侧,进一步支持将阳极电极置于臀肌中就足够了,并且可以避免手术区域出现尖锐的接地针。