Magn Reson Imaging. 2022 Nov;93:145-148. doi: 10.1016/j.mri.2022.08.010. Epub 2022 Aug 16.
To investigate the value of contrast-enhanced motion-sensitized driven equilibrium (MSDE) based black blood three-dimensional (3D) turbo spin echo (TSE) sequence in the detection of brain metastases compared with 3D Turbo Field Echo (TFE) sequence.
53 patients with suspected brain metastases were included in this study between November 2021 and February 2022. Contrast-enhanced cranial 3D TFE and MSDE-based 3D black blood TSE MR imaging were performed for each patient. Two senior neuroradiologists independently evaluated all contrast-enhanced 3D TFE and 3D black blood TSE images to detect brain metastases. The images were divided into two groups: the TFE group and the black blood TSE group. Agreement between the two reviewers for detection of the brain metastases in each group was performed using the kappa test. The two reviewers determined the final result for brain metastasis in the two groups by consensus. A paired t-test was performed for the final detection of brain metastases between the black blood TSE group and the TFE group.
There was a very good agreement between the two reviewers for the TSE group (kappa = 0.823) and a good agreement for the TFE group (kappa = 0.663). There was a statistical difference in the detection of small cortical and subcortical metastases between the TFE and the black blood TSE groups (t = 5.039, P = 0.000 < 0.05). There was no statistical difference in the detection of small supratentorial deep lesions and subtentorial lesions between the two groups.
Compared with conventional 3D TFE sequence, MSDE-based black blood 3D TSE sequence was superior for visualizing small brain metastases, especially small cortical and subcortical metastases.
探讨基于对比增强运动敏感驱动平衡(MSDE)的黑血三维(3D)快速自旋回波(TSE)序列在检测脑转移瘤方面相对于3D快速场回波(TFE)序列的价值。
2021年11月至2022年2月期间,本研究纳入了53例疑似脑转移瘤患者。对每位患者进行了对比增强头颅3D TFE和基于MSDE的3D黑血TSE磁共振成像检查。两名资深神经放射科医生独立评估所有对比增强的3D TFE和3D黑血TSE图像以检测脑转移瘤。图像分为两组:TFE组和黑血TSE组。使用kappa检验评估两位阅片者在每组中检测脑转移瘤的一致性。两位阅片者通过协商确定两组中脑转移瘤的最终结果。对黑血TSE组和TFE组之间脑转移瘤的最终检测结果进行配对t检验。
两位阅片者在TSE组的一致性非常好(kappa = 0.823),在TFE组的一致性良好(kappa = 0.663)。TFE组和黑血TSE组在检测小的皮质和皮质下转移瘤方面存在统计学差异(t = 5.039,P = 0.000 < 0.05)。两组在检测幕上小的深部病变和幕下病变方面无统计学差异。
与传统的3D TFE序列相比,基于MSDE的黑血3D TSE序列在显示小的脑转移瘤,尤其是小的皮质和皮质下转移瘤方面更具优势。