Miller G J, Vander Griend R A, Blake W P, Springfield D S
Clin Orthop Relat Res. 1987 Aug(221):246-54.
To improve intramedullary fixation with polymethylmethacrylate for use in pathologic bone, a new fixation system was developed. A fluted intramedullary rod was modified to allow injection of low-viscosity polymethylmethacrylate (PMMA) through the rod with retrograde filling of the void spaces between the rod and the medullary canal. This provided uniform distribution of the cement along the length of the rod and significantly improved fixation stability. Biomechanical testing using this system on simulated pathologic lesions of the femur showed a twofold increase in torsional strength over intramedullary methods previously used, with strength approaching that of the intact femur. In 16 patients, the device made mobilization possible immediately after the operation.
为改进聚甲基丙烯酸甲酯在病理性骨中的髓内固定效果,研发了一种新型固定系统。对有凹槽的髓内棒进行了改良,使其能够通过该棒注入低粘度聚甲基丙烯酸甲酯(PMMA),并对棒与髓腔之间的空隙进行逆行填充。这使得骨水泥沿棒的长度均匀分布,并显著提高了固定稳定性。在模拟股骨病理性病变上使用该系统进行的生物力学测试表明,与先前使用的髓内方法相比,扭转强度提高了两倍,强度接近完整股骨。在16例患者中,该装置使患者术后即可立即活动。