Suppr超能文献

在压力支持通气过程中,根据波形指导切换呼气可以改善吸气和呼气时患者与呼吸机的同步性。

Waveforms-guided cycling-off during pressure support ventilation improves both inspiratory and expiratory patient-ventilator synchronisation.

机构信息

Anaesthesia and Intensive Care, Emergency Department, Fondazione IRCCS Policlinico S. Matteo, Viale Golgi 19, 27100, Pavia, Italy; Anaesthesia, Intensive Care and Pain Therapy, Department of Clinical - Surgical, Diagnostic and Paediatric Sciences, University of Pavia, Pavia, Italy.

Anaesthesia and Intensive Care, Emergency Department, Fondazione IRCCS Policlinico S. Matteo, Viale Golgi 19, 27100, Pavia, Italy.

出版信息

Anaesth Crit Care Pain Med. 2022 Dec;41(6):101153. doi: 10.1016/j.accpm.2022.101153. Epub 2022 Sep 6.

Abstract

OBJECTIVE

To test the performance of a software able to control mechanical ventilator cycling-off by means of automatic, real-time analysis of ventilator waveforms during pressure support ventilation.

DESIGN

Prospective randomised crossover study.

SETTING

University Intensive Care Unit.

PATIENTS

Fifteen difficult-to-wean patients under pressure support ventilation.

INTERVENTIONS

Patients were ventilated using a G5 ventilator (Hamilton Medical, Bonaduz, Switzerland) with three different cycling-off settings: standard (expiratory trigger sensitivity set at 25% of peak inspiratory flow), optimised by an expert clinician and automated; the last two settings were tested at baseline pressure support and after a 50% increase in pressure support.

MEASUREMENTS AND MAIN RESULTS

Ventilator waveforms were recorded and analysed by four physicians experts in waveforms analysis. Major and minor asynchronies were detected and total asynchrony time computed. Automation compared to standard setting reduced cycling delay from 407 ms [257-567] to 59 ms [22-111] and ineffective efforts from 12.5% [3.4-46.4] to 2.8% [1.9-4.6]) at baseline support (p < 0.001); expert optimisation performed similarly. At high support both cycling delay and ineffective efforts increased, mainly in the case of expert setting, with the need of reoptimisation of expiratory trigger sensitivity. At baseline support, asynchrony time decreased from 39.9% [27.4-58.7] with standard setting to 32% [22.3-39.4] with expert optimisation (p < 0.01) and to 24.4% [19.6-32.5] with automation (p < 0.001). Both at baseline and at high support, asynchrony time was lower with automation than with expert setting.

CONCLUSIONS

Cycling-off guided by automated real-time waveforms analysis seems a reliable solution to improve synchronisation in difficult-to-wean patients under pressure support ventilation.

摘要

目的

测试一种软件在压力支持通气时通过自动实时分析呼吸机波形来控制机械通气切换的性能。

设计

前瞻性随机交叉研究。

地点

大学重症监护病房。

患者

15 名接受压力支持通气的难以撤机患者。

干预措施

使用 G5 呼吸机(Hamilton Medical,Bonaduz,瑞士)为患者进行通气,有三种不同的切换关闭设置:标准(呼气触发灵敏度设置为吸气峰流量的 25%)、由专家临床医生优化和自动化;后两种设置在基线压力支持和压力支持增加 50%时进行测试。

测量和主要结果

记录呼吸机波形并由四名精通波形分析的医生专家进行分析。检测主要和次要不同步,并计算总不同步时间。与标准设置相比,自动化将切换延迟从 407ms[257-567]减少到 59ms[22-111],无效努力从 12.5%[3.4-46.4]减少到 2.8%[1.9-4.6])在基线支持时(p<0.001);专家优化效果类似。在高支持下,两种情况下的切换延迟和无效努力都增加了,特别是在专家设置的情况下,需要重新优化呼气触发灵敏度。在基线支持时,标准设置的不同步时间从 39.9%[27.4-58.7]减少到专家优化的 32%[22.3-39.4](p<0.01)和自动化的 24.4%[19.6-32.5](p<0.001)。在基线和高支持时,自动化的不同步时间都低于专家设置。

结论

通过自动实时波形分析指导的切换关闭似乎是一种可靠的解决方案,可以改善压力支持通气下难以撤机患者的同步性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验