Department of Ophthalmology, RWTH Aachen University, Pauwelsstr. 30, 52074, Aachen, Germany.
Doc Ophthalmol. 2022 Dec;145(3):247-262. doi: 10.1007/s10633-022-09897-5. Epub 2022 Sep 10.
Visual evoked potentials (VEP) present an important diagnostic tool in various ophthalmologic and neurologic diseases. Quantitative response data varied among patients but are also dependent on the recording and stimulating equipment. We established VEP reference values for our setting which was recently modified by using a curved OLED display as visual stimulator. Distinction is made between fullfield (FF) and extrafoveal (EF) conduction, and the effect of sex, age and lens status was determined.
This prospective cross-sectional study included 162 healthy eyes of 162 test persons older than 10 years. A fullfield pattern-reversal visual evoked potential (FF-PR-VEP) with two stimulus sizes (ss) (20.4' and 1.4°) as well as an extrafoveal pattern onset-offset VEP (EF-P-ON/OFF-VEP) (ss 1.4° and 2.8°) was derived in accordance with the International Society for Clinical Electrophysiology of Vision guidelines. Amplitudes and latencies were recorded, and the mean values as well as standard deviations were calculated. Age- and sex-dependent influences and the difference between phakic and pseudophakic eyes were examined. A subanalysis of EF-P-ON/OFF-VEP and fullfield pattern onset-offset VEP (FF-P-ON/OFF-VEP) was performed. A 55-inch curved OLED display (LG55EC930V, LG Electronics Inc., Seoul, South Korea) was used as visual stimulator.
Mean P100 latency of the FF-PR-VEP was 103.81 ± 7.77 ms (ss 20.4') and 102.58 ± 7.26 ms (ss 1.4°), and mean C2 latency of the EF-P-ON/OFF-VEP was 102.95 ± 11.84 ms (ss 1.4°) and 113.58 ± 9.87 ms (ss 2.8°). For all stimulation settings (FF-PR-VEP, EF-P-ON/OFF-VEP), a significant effect of age with longer latencies and smaller amplitudes in older subjects and higher amplitudes in women was observed. We saw no significant difference in latency or amplitude between phakic and pseudophakic eyes and between EF-P-ON/OFF-VEP and FF-P-ON/OFF-VEP.
A curved OLED visual stimulator is well suited to obtain VEP response curves with a reasonable interindividual variability. We found significant effects of age and gender in our responses but no effect of the lens status. EF-P-ON/OFF-VEP tends to show smaller amplitudes.
视觉诱发电位(VEP)在各种眼科和神经科疾病中是一种重要的诊断工具。患者的定量反应数据各不相同,但也取决于记录和刺激设备。我们使用曲面 OLED 显示器作为视觉刺激器,为我们的环境建立了 VEP 参考值,该环境最近进行了修改。区分全视野(FF)和超视野(EF)传导,并确定了性别、年龄和晶状体状态的影响。
这项前瞻性的横断面研究包括 162 名年龄大于 10 岁的测试者的 162 只健康眼睛。根据国际临床视觉电生理学协会的指南,得出全视野模式反转视觉诱发电位(FF-PR-VEP)和超视野模式启动-关闭视觉诱发电位(EF-P-ON/OFF-VEP)的反应曲线,刺激大小有两种(ss)(20.4'和 1.4°)。记录振幅和潜伏期,并计算平均值和标准差。检查年龄和性别依赖性影响以及有晶状体眼和无晶状体眼之间的差异。对 EF-P-ON/OFF-VEP 和全视野模式启动-关闭视觉诱发电位(FF-P-ON/OFF-VEP)进行了亚分析。使用 55 英寸曲面 OLED 显示器(LG55EC930V,LG 电子公司,韩国首尔)作为视觉刺激器。
FF-PR-VEP 的 P100 潜伏期平均值为 103.81±7.77ms(ss 20.4')和 102.58±7.26ms(ss 1.4°),EF-P-ON/OFF-VEP 的 C2 潜伏期平均值为 102.95±11.84ms(ss 1.4°)和 113.58±9.87ms(ss 2.8°)。对于所有刺激设置(FF-PR-VEP、EF-P-ON/OFF-VEP),年龄均有显著影响,表现为年龄较大的受试者潜伏期较长、振幅较小,而女性的振幅较高。我们没有观察到有晶状体眼和无晶状体眼之间以及 EF-P-ON/OFF-VEP 和 FF-P-ON/OFF-VEP 之间的潜伏期或振幅有显著差异。
曲面 OLED 视觉刺激器非常适合获得具有合理个体间可变性的 VEP 反应曲线。我们在反应中发现了年龄和性别有显著影响,但晶状体状态没有影响。EF-P-ON/OFF-VEP 倾向于显示较小的振幅。