A study was made to expedite ion chromatography method using IonPac analytical column and self-regenerating anion suppressor for phytic acid determination in groundnut seeds and compared with a widely adopted spectrophotometric method based on enzymatic hydrolysis. The Ion Chromatography method equipped with AG11 guard and AS11 analytical columns in isocratic mode using 65 mM NaOH mobile phase at 1 mL min flow rate showed a sharp peak for phytic acid with a retention time of 2.42 ± 0.2 min. The peak area was plotted v/s concentration showed linearity with an R value of 0.997, detection limit of 0.028 mg L and recovery of 98% as against R value of 0.988 and detection limit of 0.065 mg L in the spectrophotometric method. The study demonstrates that Ion Chromatography method was more accurate with a better detection limit than spectrophotometry. Also, this method provides robust handling with lesser reagent requirements due to combined eluent generation and self-regenerating suppression.
The online version contains supplementary material available at 10.1007/s13197-022-05527-9.
进行了一项研究,以加快使用IonPac分析柱和自再生阴离子抑制器的离子色谱法测定花生种子中的植酸,并与一种广泛采用的基于酶水解的分光光度法进行比较。配备AG11保护柱和AS11分析柱的离子色谱法,在等度模式下使用65 mM NaOH流动相,流速为1 mL/min,植酸呈现出尖锐的峰,保留时间为2.42±0.2分钟。绘制峰面积与浓度的关系图,显示线性关系,R值为0.997,检测限为0.028 mg/L,回收率为98%;而分光光度法的R值为0.988,检测限为0.065 mg/L。该研究表明,离子色谱法比分光光度法更准确,检测限更好。此外,由于采用了组合洗脱液生成和自再生抑制技术,该方法操作简便,试剂需求较少。