Izumi Yuichiro, Ikeda Shuji, Kitagawa Akira, Imaeda Yusuke, Yamada Tetsuya, Hagihara Makiyo, Ota Toyohiro, Ishiguchi Tsuneo, Suzuki Kojiro
Department of Radiology, Aichi Medical University, Japan.
Department of Vascular Surgery, Aichi Medical University, Japan.
Interv Radiol (Higashimatsuyama). 2020 Sep 30;5(3):145-149. doi: 10.22575/interventionalradiology.2020-0018. eCollection 2020 Oct 30.
Percutaneous retrieval of an intravascular foreign body is a minimally invasive technique. Using cone-beam computed tomography and the lateral grasp technique, we successfully retrieved a pigtail catheter straightener that had been misinserted into the right common iliac artery. Some examples of catheter straightener retrieval have been reported; however, it is important to take care not to accidentally insert a catheter straightener into a vessel via an angiographic sheath.