基于生物物理评估,对两种具有不同黏附系统的脉搏血氧饱和度传感器在健康人体志愿者中的客观比较。

An objective comparison of two pulse oximetry sensors with different adhesive systems on healthy human volunteers based on biophysical assessments.

机构信息

cyberDERM, Inc., Broomall, Pennsylvania, USA.

Dermico LLC, Broomall, Pennsylvania, USA.

出版信息

Skin Res Technol. 2023 Jan;29(1):e13212. doi: 10.1111/srt.13212. Epub 2022 Nov 3.

Abstract

BACKGROUND

Medical Adhesive Related Skin Injuries can arise from topically applied medical devices, especially in those with fragile skin, including the elderly and premature infants. The purpose of this study was to compare gentleness and reapplication of two pulse oximetry sensors (OxySoftN and MaxN, Medtronic, Boulder, CO).

MATERIALS AND METHODS

Eighteen healthy subjects aged 65 years and older were enrolled in the gentleness trial, and 20 healthy subjects (18-69 years) were enrolled in the reapplication trial. For the gentleness trial, trans-epidermal water loss (TEWL) measurements were made at five sites on each forearm at three time points (baseline [T0], 4-h postinitial wear [T1], 4-h postsecond wear [T2]). Total amount of protein adhered to each device was also determined. For the reapplication trial, a series of 180° peel tests were performed to observe the forces required to detach the sensor from the skin.

RESULTS

TEWL rates in the tail region were significantly greater with MaxN compared to OxySoftN at T1 (p < 0.05). Both were significantly greater than control (p < 0.05). Further, protein analysis revealed that the amount of protein removed was significantly less with OxySoftN compared to MaxN (p < < 0.0001). Differences in loss of adhesion of the tail region between the two sensors were demonstrated, with OxySoftN depreciating at a much slower rate compared with MaxN.

CONCLUSION

The OxySoftN sensor appears to be gentle, even on fragile skin, based on reduced strain on the skin during removal. Further, it demonstrated the ability to withstand several reapplications without functional loss in adhesion.

摘要

背景

医用粘性相关皮肤损伤可能由局部应用的医疗设备引起,尤其是在皮肤脆弱的人群中,包括老年人和早产儿。本研究旨在比较两种脉搏血氧饱和度传感器(OxySoftN 和 MaxN,美敦力,博尔德,CO)的轻柔度和再应用性能。

材料和方法

18 名年龄在 65 岁及以上的健康受试者参加了轻柔度试验,20 名健康受试者(18-69 岁)参加了再应用试验。在轻柔度试验中,在每个前臂的五个部位进行经皮水分丢失(TEWL)测量,在三个时间点(基线[T0]、初次佩戴后 4 小时[T1]、第二次佩戴后 4 小时[T2])。还确定了每个设备上附着的总蛋白量。在再应用试验中,进行了一系列 180°剥离试验,以观察从皮肤分离传感器所需的力。

结果

与 OxySoftN 相比,MaxN 在 T1 时尾区的 TEWL 率明显更高(p<0.05)。两者均明显高于对照组(p<0.05)。此外,蛋白分析表明,OxySoftN 去除的蛋白量明显少于 MaxN(p<0.0001)。两种传感器尾区粘附丢失的差异表明,OxySoftN 比 MaxN 降解速度慢得多。

结论

OxySoftN 传感器似乎很轻柔,即使在脆弱的皮肤上也是如此,这是因为在移除过程中对皮肤的应变较小。此外,它还表现出能够承受多次重新应用而不丧失粘附功能的能力。