Analytical Enabling Capabilities, Analytical Research and Development, Merck & Co., Inc., Kenilworth, NJ 07033, United States.
Device Development, Pharmaceutical Sciences and Clinical Supply, Merck & Co., Inc., Rahway, NJ 07065, United States.
J Pharm Sci. 2022 Dec;111(12):3191-3194. doi: 10.1016/j.xphs.2022.09.015. Epub 2022 Sep 19.
Silicone oil is often applied to the inner surface of glass syringes and cartridges to reduce friction between the glass surface and elastomeric plunger stopper. This oil can appear as intrinsic and non-proteinaceous particles in the ejected fluid or drug product. Limited data is available to understand the impact of age (time between syringe manufacture and filling) on silicone oil migration into the drug product. This study compares subvisible particle count and extrusion force of siliconized syringes from two different manufacturers stored at ambient condition for 2-3 (fresh syringes) and 13-14 (aged syringes) months then filled and placed at 40°C for an additional three months. The fresh syringes exhibit a 2.5-fold increase in subvisible particle count compared to those aged ones. Moreover, the fresh syringes exhibit up to a 2-fold increase in extrusion force. These findings suggest the degree and amount of silicone oil migration is influenced by the time in storage of the glass syringe prior to filling. This rapid communication highlights syringe storage time prior to filling as a factor to be considered during development.
硅油通常涂在内置玻璃注射器和药筒的内表面,以减少玻璃表面和弹性柱塞密封件之间的摩擦。这种油可能会以固有且非蛋白性颗粒的形式出现在喷出的液体或药物产品中。目前可用的数据有限,无法了解注射器制造和填充之间的时间(老化时间)对硅油迁移到药物产品中的影响。本研究比较了来自两个不同制造商的硅化注射器的亚可见颗粒计数和挤出力,这些注射器在环境条件下储存了 2-3 个月(新鲜注射器)和 13-14 个月(老化注射器),然后填充并在 40°C 下再放置三个月。与老化的注射器相比,新鲜的注射器中亚可见颗粒计数增加了 2.5 倍。此外,新鲜的注射器的挤出力增加了 2 倍。这些发现表明,在填充之前玻璃注射器的储存时间会影响硅油的迁移程度和数量。这一快速交流强调了在填充之前的注射器储存时间是在开发过程中需要考虑的一个因素。