Suppr超能文献

一例因激光导线拔除导致中心静脉导管断裂的独特病例。

A Unique Case of Central Line Fracture Due to Laser Lead Extraction.

作者信息

Towheed Arooge, Smith Alexander, Ghawanmeh Malik, Tsimploulis Apostolos, Hadadi Cyrus, Thomaides Athanasios

机构信息

Department of Electrophysiology, MedStar Washington Hospital Center, Washington, DC, USA.

Department of Cardiology, MedStar Washington Hospital Center, Washington, DC, USA.

出版信息

JACC Case Rep. 2022 Nov 3;4(24):101633. doi: 10.1016/j.jaccas.2022.08.044. eCollection 2022 Dec 21.

Abstract

Transvenous laser-assisted lead extraction is successful, with a low procedural complication rate for a wide range of indications. Here, we report a case of right internal jugular triple-lumen central venous catheter fracture and subsequent embolism to the right pulmonary artery during laser lead extraction that was successfully retrieved with a gooseneck snare. ().

摘要

经静脉激光辅助导线拔除术成功率高,适用于多种适应症,手术并发症发生率低。在此,我们报告一例右颈内静脉三腔中心静脉导管断裂并在激光导线拔除过程中随后栓塞至右肺动脉的病例,该导管通过鹅颈圈套器成功取出。()